Hypertension. 1995;26:835-842

This Article
Right arrow Abstract Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by O’Brien, E.
Right arrow Articles by Staessen, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by O’Brien, E.
Right arrow Articles by Staessen, J.

(Hypertension. 1995;26:835.)
© 1995 American Heart Association, Inc.


Articles

State of the Market

A Review of Ambulatory Blood Pressure Monitoring Devices

Eoin O’Brien; Neil Atkins; Jan Staessen

From The Blood Pressure Unit, Beaumont Hospital (E.O’B., N.A.), Dublin, Ireland, and Klinisch Laboratorium Hypertensie, Inwendige Geneeskunde-Cardiologie (J.S.), UZ Gasthuisberg, Leuven, Belgium.


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowABPM Devices Not Subjected...
down arrowABPM Devices Validated by...
down arrowABPM Devices Validated According...
down arrowABPM Devices Validated According...
down arrowDiscussion
down arrowAppendix 1
down arrowReferences
 
Abstract The introduction of 24-hour ambulatory blood pressure measurement into clinical practice created a large market for ambulatory blood pressure measurement devices. Forty-three such devices from 31 manufacturers or suppliers are now available to satisfy a market demand that is likely to increase. The aim of this article is to identify the devices available and then to examine critically any validation studies assessing accuracy and performance. Of the 43 devices available 18 have been validated according to the protocols of the Association for the Advancement of Medical Instrumentation (AAMI) or the British Hypertension Society (BHS) in 25 reported studies. In 9 of these studies the protocol was not adhered to, and the results, which are therefore questionable, are noted but not considered further. Fourteen devices were evaluated according to the accuracy criteria of both protocols, and of these 9 fulfilled the requirements. From this review of 43 devices on the market it may be concluded that, at the time of writing, there is published evidence for only 9 devices meeting the generally accepted AAMI and BHS criteria for accuracy and performance; these are the A&D TM-2420 models 6 and 7 and TM-2421, CH-Druck, Nissei ABPM DS-240, Profilomat, QuietTrak, and SpaceLabs SL-90202 and SL-90207.


Key Words: blood pressure monitoring, ambulatory • equipment design • evaluation studies • durable medical equipment


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowABPM Devices Not Subjected...
down arrowABPM Devices Validated by...
down arrowABPM Devices Validated According...
down arrowABPM Devices Validated According...
down arrowDiscussion
down arrowAppendix 1
down arrowReferences
 
When the technique of blood pressure measurement was introduced to clinical medicine in the early years of the 20th century the importance of accuracy and the limitations of the technique were well recognized.1 However the standards called for by the clinicians and scientists who pioneered the technique were relaxed as the 20th century progressed. Now once again the methodology of measurement in both clinical practice and research in hypertension is recognized as a cause for concern.2 The first serious approach to device validation was that of Halls Dally,3 who in 1926 called in the help of The National Physics Laboratory to endorse the claim of the manufacturer regarding the accuracy of the then-new Baumanometer. Since the 1930s various national bodies such as the American Heart Association, the British Cardiac Society, and the World Health Organization have endorsed the importance of the accuracy of devices, but none have stated how this was to be achieved.4 5 6 7 8 9 10 11 12 Likewise national standards institutions have emphasized the issue but usually do not have the authority to impose minimal accuracy standards.13 14 15 The importance of accurate devices recently has been voiced more strongly by individuals involved in hypertension research, as is evident from the growing number of publications on this subject.16

In the 1960s and 1970s individual groups, frustrated by the failure of manufacturers to produce evidence to match their claims, began to validate measuring systems according to a variety of protocols and so illustrated the need for independent device validation.17 18 19 20 21 22 23 24 25 Several of these ad hoc protocols have been applied to the evaluation of some of the more expensive ABPM systems.26 27 28 29 30 31 However well intentioned such protocols may be, they have the serious disadvantage of not permitting comparison between devices because of the differing methods of validation in these protocols.32

In 1987 the AAMI published the American National Standard for Electronic or Automated Sphygmomanometers, which included a protocol for the evaluation of device accuracy,33 and publication of this protocol was followed in 1990 by publication of the protocol of the BHS.34 These protocols, which differed in detail, had a common objective, namely to standardize validation methods and thus to establish minimum standards of accuracy and performance as well as facilitate comparison between devices.35 36 Both protocols have been revised recently.37 38 39 40 Although other countries such as Germany and Australia have included recommendations for testing the accuracy of ABPM devices in their national standards, validation studies have not been published for these protocols, which therefore will not be considered further in this review. Ng and Small41 42 have reviewed the differences between national protocols for validation in considerable detail.

Some workers have protested protocol standardization, however, being of the view that innovative methods of device evaluation might be discouraged.43 In particular, the BHS reservations concerning comparison of noninvasive devices with direct intra-arterial measurement have been questioned.43 44 These reservations not only were based on the ethical issue of performing intra-arterial measurement in healthy volunteers but also arose because intra-arterial techniques give different blood pressures than those obtained by noninvasive methods.40 Nevertheless, the revised BHS protocol recognized that centers with expertise in intra-arterial measurement might be able to provide valuable information on performance of devices in ambulatory conditions, which otherwise might not be obtained.40

It is therefore possible to identify five different approaches to categorizing ABPM devices for the development of validation procedures: (1) devices that have not been validated independently; (2) devices that have been validated by ad hoc protocols; (3) devices that have been validated according to the AAMI protocol alone; (4) devices that have been validated according to the BHS and AAMI protocols; and (5) devices that have been validated by use of intra-arterial comparison methods. The first four approaches are the subject of this review. Intra-arterial comparisons are difficult to evaluate because of differing methodologies, and their main function has been to provide information on performance during exertion.45 46 47 48 49 50 51 52 53


*    ABPM Devices Not Subjected to Validation According to AAMI or BHS Protocols
up arrowTop
up arrowAbstract
up arrowIntroduction
*ABPM Devices Not Subjected...
down arrowABPM Devices Validated by...
down arrowABPM Devices Validated According...
down arrowABPM Devices Validated According...
down arrowDiscussion
down arrowAppendix 1
down arrowReferences
 
In a recent review of more than 400 automated noninvasive blood pressure monitors Ng and Small54 55 reviewed the methods of ABPM devices currently available. By combining data from the report of Ng and Small with our research we have identified 43 ABPM systems (from 31 suppliers and manufacturers; see "Appendix") that have been marketed in recent years and that could be in clinical use at the time of writing (Table 1). We estimate that 10 of these systems have been superseded by later models but may still be in clinical use.


View this table:
[in this window]
[in a new window]
 
Table 1. Forty-three ABPM Systems Currently on the Market


View this table:
[in this window]
[in a new window]
 
Table 1A. (Continued)

The BHS protocol recommends that the results of validation studies be published in peer-reviewed journals, and a search of the literature shows that of the 43 systems identified only 18 have been subjected to independent validation with either the BHS or AAMI protocol. Put another way only 9 of the 31 suppliers and manufacturers have had their products validated (Table 1). Our listing of those devices that have not been independently validated does not carry any implications as to their accuracy but merely informs readers that to date such validations have not been performed or if they have the results have not been published in the indexed medical journals. We also acknowledge that many of the ABPM devices now available have been produced only recently, and procedures for validations may be planned or under way.


*    ABPM Devices Validated by Ad Hoc Protocols
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowABPM Devices Not Subjected...
*ABPM Devices Validated by...
down arrowABPM Devices Validated According...
down arrowABPM Devices Validated According...
down arrowDiscussion
down arrowAppendix 1
down arrowReferences
 
Validations according to various ad hoc protocols were the first step in evaluation of ABPM and other blood pressure measurement systems.17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Although many useful observations were made in these studies they are now of historical interest rather than being applicable to contemporary practice because most of the devices so validated have been replaced by newer models. However, these studies formed the foundation on which the AAMI and BHS built their standardized protocols.38 39


*    ABPM Devices Validated According to the AAMI Protocol Alone
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowABPM Devices Not Subjected...
up arrowABPM Devices Validated by...
*ABPM Devices Validated According...
down arrowABPM Devices Validated According...
down arrowDiscussion
down arrowAppendix 1
down arrowReferences
 
The ABPM systems of three manufacturers have been validated according to the first AAMI protocol33 : the Accutracker, the Medilog, and the TM-2420 (Table 2). Nine validation studies were performed on various models of these devices, but all nine studies failed to adhere strictly to the AAMI protocol recommendations.52 56 57 58 59 60 61 62 63 An important protocol violation that occurred in eight studies was substitution of the Hawksley random-zero sphygmomanometer for the mercury sphygmomanometer recommended in the protocol52 56 57 58 59 61 62 63 to reduce observer bias, but the Hawksley device subsequently was shown to underestimate blood pressure64 ; we previously presented the consequences that this may have on validation studies.65 In a further study we compared a database of paired blood pressure measurements made with the Hawksley random-zero sphygmomanometer and a standard mercury sphygmomanometer with a database of paired measurements made with a SpaceLabs SL-90202 ambulatory recorder and a standard sphygmomanometer to determine how the SL-90202 recorder would have fared if it had been assessed against the Hawksley random-zero sphygmomanometer instead of a standard sphygmomanometer. Replacing the standard with the Hawksley sphygmomanometer reversed the direction of the average measurement error found and demoted the SL-90202 recorder measurement from BHS grades C and B, for systolic and diastolic accuracy, respectively, to grade D overall, the lowest rating of accuracy in the BHS grading system.66 Thus the conclusions of the eight validation studies in which the Hawksley device was substituted for the standard sphygmomanometer are questionable52 56 57 58 59 61 62 63 because we do not know how the results of these studies were affected by the use of the Hawksley sphygmomanometer.


View this table:
[in this window]
[in a new window]
 
Table 2. Results of Nine Validation Studies Performed With the AAMI Protocol Alone

Another serious protocol violation, which occurred in two studies, was that the numbers of subjects recruited were well below that stipulated in the AAMI and BHS protocols, and thus the results also must be regarded as questionable.57 60

Another confounding issue apart from the importance of adhering strictly to protocols in conducting validation studies is the lack of information as to which device model is being validated. The BHS protocols emphasize the importance of having the manufacturer indicate any modifications made to ABPM devices by changing the model number.34 40 The importance of this stricture is well illustrated by the conflicting reports from several laboratories on the accuracy of the Takeda TM-2420 (A&D).52 60 61 62 63 67 68 The results of individual studies on this device, which have been reviewed in detail elsewhere,69 show that apparent differences between laboratories can be accounted for by the manufacturers submitting a different model for validation without indicating whether the original device was modified. We hope this trend has passed, and it is perhaps significant that recent reports on the A&D systems stipulate which model is being used.49 70 71


*    ABPM Devices Validated According to AAMI and BHS Protocols
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowABPM Devices Not Subjected...
up arrowABPM Devices Validated by...
up arrowABPM Devices Validated According...
*ABPM Devices Validated According...
down arrowDiscussion
down arrowAppendix 1
down arrowReferences
 
Thirteen ABPM systems70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 have been evaluated according to both the original BHS34 and AAMI criteria,33 and more recently the revised AAMI protocol was used for validation (Table 3).81 These studies, unlike the earlier studies mentioned above in which only the AAMI protocol was used, complied generally with BHS and AAMI protocol requirements. Moreover, two validation studies in which both protocols were used have been performed with pregnant women.83 84


View this table:
[in this window]
[in a new window]
 
Table 3. ABPM Systems Subjected to Validation According to the BHS and AAMI Protocols

Expert committees38 40 and steering committees of some studies85 86 87 have recommended independent validation of devices for clinical practice and research. Of the many ABPM devices available, only nine fulfill both the BHS and AAMI criteria in that they achieved at least a grade of B/B for both systolic and diastolic pressures,40 and the mean difference between the ambulatory device and a mercury standard was less than 5 mm Hg with a standard deviation of less than 8 mm Hg.38 These devices are the CH-Druck, Nissei ABPM DS-240, Profilomat, QuietTrak, SpaceLabs SL-90202 and SL-90207, and A&D Takeda TM-2420 models 6 and 7 and TM-2421. Moreover, the SpaceLabs SL-90207 fulfilled the criteria of both protocols in pregnant women83 84 (Table 4).


View this table:
[in this window]
[in a new window]
 
Table 4. Nine ABPM Devices That Have Fulfilled BHS and AAMI Accuracy Criteria as of January 1995

Two devices, the Novacor DIASYS 200 R75 and the Accutracker II (version 30/23),80 fulfilled the AAMI criteria for systolic and diastolic pressures and satisfied the BHS criteria for systolic but not diastolic pressure. The early A&D TM-2420 models72 and the Del Mar Avionics Pressurometer IV76 failed to fulfill the criteria of either protocol.


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowABPM Devices Not Subjected...
up arrowABPM Devices Validated by...
up arrowABPM Devices Validated According...
up arrowABPM Devices Validated According...
*Discussion
down arrowAppendix 1
down arrowReferences
 
ABPM is now recognized to be a useful procedure in clinical practice, an occurrence that has created a large market with many ABPM device manufacturers. The potential purchaser faced with attractive advertising brochures and persuasive sales talk may have difficulty determining the accuracy of any given ABPM device. Two national bodies, the AAMI38 and the BHS,40 have recommended that all blood pressure measuring devices be validated carefully before being introduced into clinical practice. This review was undertaken to examine the published evidence for the accuracy and performance of all known ambulatory systems on the market.

We identified 43 ABPM devices from 31 suppliers and manufacturers. Three devices had been validated by the AAMI protocol alone in nine studies, but the protocol had been violated in all instances. The results are therefore questionable, and we did not consider these devices in the final assessment. Fourteen ABPM devices have been evaluated according to both protocols. Of these, nine fulfilled both the BHS and AAMI criteria in that they achieved at least a grade of B/B for systolic and diastolic pressures and the mean difference between the ambulatory device and a mercury standard was less than 5 mm Hg with a standard deviation less than 8 mm Hg.38 40

As noted above, of 43 ABPM devices available only 9 fulfilled AAMI and BHS criteria to ensure reasonable accuracy in clinical use (Table 4), and for 28 devices no published evidence of independent validation according to either the AAMI or BHS protocol could be found at the time this article was written. We again emphasize that the publishing process is often lengthy and that manufacturers may have conducted or may be conducting independent validation studies on many of the devices listed in this review. However we also emphasize the importance of publishing such results as recommended by the BHS protocol40 so that device accuracy and performance occasionally can be reviewed. One might argue that it should be sufficient for the manufacturer or supplier to provide evidence of independent validation according to a standardized, albeit unpublished, protocol, but the evidence from this review demonstrating that protocol violations are common also illustrates the need for the critical peer-review process that is a prerequisite of scientific publication. One problem with this policy is that the hypertension journals have scarce space to provide for publication of somewhat repetitive, albeit important, data. The Journal of Hypertension facilitated the early publication of validation studies by publishing short summary reports of validation studies within the regular journal issue and then publishing the full report later in a supplement. The Journal of Blood Pressure Monitoring, which is to be launched soon, will provide for publication of validation studies among a number of other functions in this growing area of interest (Prof William White, The University of Connecticut Health Center, Farmington, personal communication, 1995).

We are aware of two untoward practices to which attention should be drawn. First, some manufacturers may have applied the favorable validation results of one model to another model that has not been validated. The BHS protocol clearly states that each model must be subjected to independent validation.40 Second, some manufacturers claim to have satisfied a protocol by quoting the results of published work in which the AAMI or BHS criteria are applied to data derived from a validation study that did not adhere to the protocol requirements.

This review of the state of the market also has allowed us to look to the trends in manufacturing practice. The next generation of devices for ABPM will be smaller, quieter, and considerably cheaper than many of its predecessors. The trend is toward oscillometric rather than auscultatory measurement techniques, although many devices may combine both methods. Other methods of measurement such as pulse-wave–velocity detection, which is not dependent on cuff occlusion, may be developed.54 The vascular unloading technique, which is used in the Finapres device88 and which gives continuous rather than intermittent blood pressures, is confined to research laboratories at present, but a clinical role can be anticipated. ECG gating is being used less frequently, and as algorithms improve its use for conferring greater accuracy will have to be balanced against the major disadvantages of the time required for its application and the discomfort to the male subject whose chest must be shaved before use. Devices that combine ECG and ABPM techniques are likely to be used in certain clinical circumstances. The combination of a motion-logging system and ABPM would be a welcome development that would permit standardization of ABPM in relation to activity, especially when repeat ABPMs are being performed.

Device validations are protracted procedures requiring considerable involvement of trained personnel and careful supervision.36 It has been estimated that the cost of performing a full validation according to the original BHS protocol is about British sterling £25 000.36 This figure is likely to be appreciably higher for validation studies performed according to the revised protocols, particularly if both the AAMI and BHS criteria are fulfilled.89 Likewise, ABPM systems incorporating two measurement methods, such as Korotkov sound detection and oscillometry, should be validated for both methods, an exercise that adds appreciably to the cost of validation. Future technological developments should improve device validation procedures. For example, noninvasive blood pressure simulators now being developed can be expected to reduce dependence on hypertensive subjects for validation of oscillometric devices.41 42 The manufacture of a reliable automated device will reduce dependency on observer-measured blood pressure so that it will be possible to simplify further the validation procedure. The introduction of audiovisual methods of recording Korotkov sounds simultaneously with the use of a falling column of mercury also may be expected to reduce dependency on observers.90 However, it must be emphasized that all such innovative techniques must be subjected to careful validation before being introduced into research methodology.

The adoption of standards by ABPM device manufacturers may not be easily effected. One of the important points to emerge from this review is the need to persuade manufacturers of the importance of having the ABPM devices validated before placing these devices on the market. In fact, this principle should apply to all blood pressure measuring devices. At present manufacturers are not obliged to guarantee the accuracy of their products, although most reputable manufacturers welcome the opportunity to have their devices evaluated independently according to a generally accepted protocol. The European community has established a working party (CEN/TC 205/WG 10 Non-invasive sphygmomanometers) to draw up a standard for all ABPM devices, and in 1996 a directive will be issued that will be legally binding on member states.91

The process of encouraging manufacturers of ABPM systems to have their products evaluated independently according to an approved evaluation procedure has been influenced positively by editors of general medical, clinical pharmacology, and hypertension journals seeking evidence supporting the accuracy of ABPM systems used in research studies. Health authorities and sponsoring organizations should purchase only equipment that has been evaluated adequately. The protocols of some hypertension studies, such as the Syst-Eur (Systolic Hypertension in the Elderly) study,85 the APTH (Ambulatory Blood Pressure and Treatment of Hypertension) study,86 and the OvA (Office Versus Ambulatory Measurement) study,87 stipulate that no automated system can be used in the study unless it is evaluated independently by use of an accepted validation protocol.

Our concluding statement is devoted to expressing a degree of scientific disquiet. The AAMI published its protocol in 198733 and revised it in 1993,38 and in 1990 the BHS constituted a Working Party that published a protocol for validation of ABPM devices34 that was revised in 1993.40 In spite of this we have had to write six letters in the year before publication of this article alerting researchers and potential purchasers of blood pressure measurement devices that the results of validation studies published in peer-reviewed journals were questionable or invalid because the devices being considered failed to comply with protocol requirements.92 93 94 95 96 97 That such letters have had to be written at all is disquieting but that they may inevitably fail to influence the published statement of the original article is a cause for even greater concern.


*    Selected Abbreviations and Acronyms
 
AAMI = Association for the Advancement of Medical Instrumentation
ABPM = ambulatory blood pressure measurement
BHS = British Hypertension Society


*    Acknowledgments
 
We acknowledge with gratitude support from the Charitable Infirmary Charitable Trust, the Royal College of Surgeons of Ireland, and Beaumont Hospital in Ireland. We thank Dr Kim-Gau Ng from the Department of Mechanical Engineering at Queen’s University, Ontario, Canada, who not only read the manuscript and offered many useful suggestions but also gave us permission to use much of his data in Table 1.


*    Footnotes
 
Reprint requests to Prof E. O’Brien, The Blood Pressure Unit, Beaumont Hospital, Dublin 9, Ireland.


*    Appendix 1
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowABPM Devices Not Subjected...
up arrowABPM Devices Validated by...
up arrowABPM Devices Validated According...
up arrowABPM Devices Validated According...
up arrowDiscussion
*Appendix 1
down arrowReferences
 
ABPM Device Suppliers and Manufacturers
A&D Co, Ltd, 3-23-14 Higashi-Ikeburo, Tokyo, Japan; phone, 3-5391-6123; fax, 3-5391-6148.

BioAnalogics Systems Inc, 9000 SW Gemini, Beaverton, OR 97005; phone, 800-327-7953; fax, 503-641-4031.

Biotrac Inc, 9261 130th Ave N, Largo, FL 34643-1304; phone, 813-584-9129; fax, 813-397-9893.

Circadian Inc, 3942 N First St, San Jose, CA 95134; phone, 800-669-7001; fax, 408-434-9585.

Colin Medical Instruments Corp, University Technology Park, 5850 Farinon Dr, San Antonio, TX 75249; phone, 800-829-6427; fax, 210-696-8808.

Del Mar Avionics, 1601 Alton Ave, Irvine, CA 92714-4878; phone, 800-854-0481; fax, 714-261-0529.

Disetronic Medical Systems AG, Brunnmattstr 6, CH-3401 Burgdorf, Switzerland; phone 41-34-23-13-33; fax, 41-34-23-13-85.

GH Medical Inc, 2010 E Hennepin Ave, Minneapolis, MN 55413; phone, 612-623-3966; fax, 612-378-1937.

Healthcare Technology Ltd, York House, City Fields Business Park, Tangmere, West Sussex, UK; phone, 44-1243-528-800; fax, 44-1243-774-728.

Hill-Med Corp, 7215 NW 46th St, Miami, FL 33166; phone, 305-594-7474; fax, 305-477-0699.

IDT France, 15 rue Boileau, 66000 Perpignan, France; phone, 33-68-35-4620; fax, 33-68-34-0945.

I.E.M. Electromedicina, S.L, Plaça del Poble, 27, 08410 Vilanova del Vallès, Barcelona, Spain; phone, 34-3-845-61-28; fax, 34-3-845-60-12.

I.E.M. GmBH, Eifelstr 20, D-52224 Stolberg, Germany; phone, 49-2402-7-59-01; fax, 49-2402-7-28-19.

Imex Medical Systems Inc, 6355 Joyce Dr, Golden, CO 80403; phone, 800-525-2519; fax, 303-431-0429.

Instromedix Inc, One Technology Center, 7431 NE Evergreen Pkwy, Hillsboro, OR 97124-5898; phone, 800-633-3361; fax, 503-681-8230.

Kontron Instruments, Via G. Fantoli, 16/15, 20138 Milano, Italy; phone, 39-2-50721; fax, 39-2-506-0918.

Novacor, 4 passage Saint-Antoine, 92508 Rueil-Malmaison, Cedex, France; phone, 33-1-47-08-06-66; fax, 33-1-47-32-45-76.

Oxford Medical, 1 Kimber Rd, Abingdon, Oxon OX14 1BZ, UK; phone, 44-1235-533-433; fax, 44-1235-534-465.

PAR Medizintechnik GmBH, Einemstr 9, D-10787 Berlin, Germany; phone, 49-30-2-13-90-55; fax, 49-30-2-13-85-42.

Pilger Medizin-Elektronik, Speicherstr 48, 9000 St Gallen, Germany; phone, 49-71-22-30-16; fax, 49-71-22-30-04.

PulseTrend Inc, 2010 E Hennepin Ave, Suite 220, Minneapolis, MN 55413; phone, 800-584-6686; fax, 612-623-7748.

Save 33 Electronique Medicale, PO Box 25, 124 rue Emile Zola, 59860 Bruay sur Escaut, France; phone, 33-27-29-55-44; fax, 33-27-29-55-67.

Schiller AG, Altgasse 68, CH-6340 Baar, Switzerland; phone, 41-42-33-43-53; fax, 41-42-31-08-80.

SpaceLabs Medical Inc, PO Box 97013, 15220 NE 40th St, Redmond, WA 98073-9713; phone, 800-251-9910; fax, 206-885-4877.

Stuart Medical Inc, 11403 Cronhill Dr, Suite B, Owings Mills, MD 21117; phone, 800-377-5378; fax, 410-581-1513.

Suntech Medical Instruments Inc, 8608 Jersey Court, Raleigh, NC 27613; phone, 800-421-8626; fax, 919-782-9113.

Suzuken Co Ltd, 1-6-10 Shimizu, Kita-ku, Nagoya 462, Japan; phone, 81-52-971-3641; fax, 81-52-962-7440.

TNO Biomedical Instrumentation, Suite G1-111, Meibergdreef 9, 1105 AZ Amsterdam, the Netherlands; phone, 20-566-58-44; fax, 20-697-64-24.

Tycos Instruments Inc, A Welch Allyn Co, 95 Old Shoals Rd, Ardern, NC 28704; phone, 704-684-4895; fax, 704-687-1002.

Zewa AG, Seestr 7a, CH-6052 Hergiswil, Switzerland; phone, 41-36-51-88; fax, 41-36-53-36.

Zymed Inc, 20 N Aviador St, CA 93010-8302; phone, 800-235-5941; fax, 805-987-9532.

Received May 2, 1995; first decision June 6, 1995; accepted August 21, 1995.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowABPM Devices Not Subjected...
up arrowABPM Devices Validated by...
up arrowABPM Devices Validated According...
up arrowABPM Devices Validated According...
up arrowDiscussion
up arrowAppendix 1
*References
 
1. O’Brien E, Fitzgerald D. The history of indirect blood pressure measurement. In: Birkenhager WH, Reid JL. Handbook of Hypertension. Amsterdam, the Netherlands: Elsevier; 1991:1-54.

2. Roche V, O’Malley K, O’Brien E. How ‘scientific’ is blood pressure measurement in leading medical journals? J Hypertens.. 1990;8:1167-1168. [Medline] [Order article via Infotrieve]

3. Halls Dally JF. In: Heinemann W. High Blood Pressure: Its Variations and Control. 2nd ed. London, UK;1926:13, 33.

4. Joint Recommendations of the American Heart Association and the Cardiac Society of Great Britain and Ireland: standardization of blood pressure readings. New York, NY: American Heart Association; 1939:7.

5. Committee to Revise Standardization of High Blood Pressure Readings. Recommendations of human blood pressure determinations by sphygmomanometers. New York, NY: American Heart Association; 1951:10.

6. Report of Expert Committee. Hypertension and coronary heart disease: classification and criteria for epidemiological studies. Geneva, Switzerland: World Health Organization; 1959:9. World Health Organization Technical Report Series No. 168.

7. Report of Expert Committee. Hypertension and ischaemic heart disease: preventive aspects. Geneva, Switzerland: World Health Organization; 1962:231;4. World Health Organization Technical Report Series.

8. American Heart Association. Recommendations for human blood pressure determination by sphygmomanometers: Central Committee for Medical and Community Program of the American Heart Association. New York, NY: American Heart Association; 1967:23.

9. Kirkendall WM, Feinleib M, Freis ED, Mark AL. Recommendations for human blood pressure determination by sphygmomanometers: subcommittee of the Postgraduate Education Committee. Circulation. 1980;62:1145A-1155A.

10. Petrie JC, O’Brien ET, Littler WA, de Swiet M. Recommendations on blood pressure measurement: British Hypertension Society. BMJ.. 1986;293:611-615.

11. Perloff D, Grim C, Flak J, Frohlich ED, Hill M, McDonald M, Morgenstern BZ. Human blood pressure determination by sphygmomanometry: American Heart Association Medical/Scientific Statement: special report. Circulation. 1993;88:2460-2470. [Free Full Text]

12. Petrie JC, O’Brien ET, Littler WA, de Swiet M. Recommendations on Blood Pressure Measurement: British Hypertension Society. London, UK: British Medical Journal Books; 1990:19.

13. Federal specification for sphygmomanometers: aneroid and mercurial. Washington, DC: Superintendent of Documents; 1978. US Government Printing Office GG-S-618D.

14. Australian Standard: sphygmomanometers. In: Standards Australia. N Sydney, New South Wales, Australia: Standards Association of Australia; 1989;(No. AS 3655-1989):36.

15. Specification for aneroid and mercury non-automated sphygmomanometers. London, UK: British Standards Institution; 1990. Revision of BS-2743 (1956) and BS-2744 (1956).

16. O’Brien E, O’Malley K. In: Birkenhager WH, Reid JL, eds. Clinical blood pressure measurement. In: Robertson JIS, ed. Handbook of Hypertension: Vol 15. Amsterdam, the Netherlands: Elsevier; 1992:14-50.

17. Labarthe DR, Hawkins CM, Remington RD. Evaluation of performance of selected devices for measuring blood pressure. Am J Cardiol.. 1973;32:546-553. [Medline] [Order article via Infotrieve]

18. Hunyor SN, Flynn JM, Cochineas C. Comparison of performance of various sphygmomanometers with intra-arterial blood pressure readings. BMJ.. 1978;2:159-162.

19. Barker WF, Hediger ML, Katz SH, Bowers EJ. Concurrent validity studies of blood pressure instrumentation: the Philadelphia Blood Pressure Project. Hypertension.. 1984;6:85-91. [Abstract/Free Full Text]

20. Sloan PJM, Zezulka A, Davies P, Sangal A, Beevers M, Beevers G. Standardised methods for comparison of sphygmomanometers. J Hypertens.. 1984;2:547-551. [Medline] [Order article via Infotrieve]

21. Steiner R, Luscher T, Boerlin H-J, Siegenthaler W, Vetter W. Clinical evaluation of semiautomatic blood pressure devices for self-recording. J Hypertens. 1985;3(suppl 1):23-25.

22. Pickering TG, Cvetkovski B, James GD. An evaluation of electronic recorders for self-monitoring of blood pressure. J Hypertens. 1986;4(suppl 5):S328-S330.

23. Hahn LP, Folsom AR, Sprafka JM, Prineas RJ. Prevalence and accuracy of home sphygmomanometers in an urban population. Am J Public Health.. 1987;77:1459-1461. [Abstract/Free Full Text]

24. Evans CD, Haynes RB, Goldsmith CH, Hewson SA. Home blood pressure-measuring devices: a comparative study of accuracy. J Hypertens.. 1989;7:133-142. [Medline] [Order article via Infotrieve]

25. Fitzgerald D, O’Callaghan W, O’Malley K, O’Brien E. Inaccuracy of the London School of Hygiene sphygmomanometer. BMJ.. 1982;284:18-19.

26. Kain HK, Hinman AT, Sokolow M. Arterial blood pressure measurements with a portable recorder in hypertensive patients, I: variability and correlation with ‘Casual’ Pressures. Circulation. 1964;30:882-892. [Free Full Text]

27. Harshfield GA, Pickering TG, Laragh JH. A validation of the Del Mar Avionics ambulatory blood pressure system. Ambul Electrocardiogr.. 1979;1:7-12.

28. Fitzgerald DJ, O’Callaghan WG, McQuaid R, O’Malley K, O’Brien E. Accuracy and reliability of two indirect ambulatory blood pressure recorders: Remler M2000 and Cardiodyne Sphygmolog. Br Heart J.. 1982;48:572-579.[Free Full Text]

29. Clement DL, De Pue N, Packet L, Bobelyn M, Van Maele GO. Performance of noninvasive ambulatory blood pressure recordings. J Hypertens. 1983;1(suppl 2):296-298.

30. Brunner HR, des Combes BJ, Waeber B, Porchet M. Accuracy and reproducibility of ambulatory blood pressure recordings obtained with the Remler system. J Hypertens. 1983;1(suppl 2):291-292.

31. Gould BA, Hornung RS, Kieso HA, Altman DA, Cashman PMM, Raftery EB. Evaluation of the Remler M2000 blood pressure recorder: comparison with intraarterial blood pressure recordings both at hospital and at home. Hypertension.. 1984;6:209-215. [Abstract/Free Full Text]

32. O’Brien E, O’Malley K, Sheridan J. The need for a standardized protocol for validating non-invasive ambulatory blood pressure measuring devices. J Hypertens. 1989;7(suppl 3):S19-S20.

33. American National Standard for Electronic or Automated Sphygmomanometers: ANSI/AAMI SP10-1987. Arlington, Va: Association for the Advancement of Medical Instrumentation (3330 Washington Boulevard, Suite 400, Arlington, VA 22201-4598); 1987:25.

34. O’Brien E, Petrie J, Littler WA, Padfield PL, O’Malley K, Jamieson M, Altman D, Bland M, Atkins N. British Hypertension Protocol: evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens.. 1990;8:607-619. [Medline] [Order article via Infotrieve]

35. O’Brien E, O’Malley K. Twenty-four-hour ambulatory blood pressure monitoring: a review of validation data. J Hypertens. 1990;8(suppl 6):S11-S16.

36. O’Brien E, O’Malley K. Evaluation of blood pressure measuring devices with special reference to ambulatory systems. J Hypertens. 1990;8(suppl 7):S133-S139.

37. White WB, Berson AS, Robbins C, Jamieson MJ, Prisant LM, Roccella E, Sheps SG. National standard for measurement of resting and ambulatory blood pressure with automated sphygmomanometers. Hypertension.. 1993;21:504-509. [Abstract/Free Full Text]

38. American National Standard for Electronic or Automated Sphygmomanometers: ANSI/AAMI SP 10-1992. Arlington, Va: Association for the Advancement of Medical Instrumentation (3330 Washington Boulevard, Suite 400, Arlington, VA 22201-4598); 1993: P 40.

39. O’Brien E, Petrie J, Littler WA, de Swiet M, Padfield PL, Altman D, Bland M, Coats A, Atkins N. Short report: an outline of the British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens.. 1993;11:677-679. [Medline] [Order article via Infotrieve]

40. O’Brien E, Petrie J, Littler WA, de Swiet M, Padfield PL, Altman D, Bland M, Coats A, Atkins N. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens. 1993;11(suppl 2):S43-S63.

41. Ng K-G, Small C. Review of methods and simulators for evaluation of noninvasive blood pressure monitors. J Clin Eng.. 1992;17:469-479. [Medline] [Order article via Infotrieve]

42. Ng K-G, Small C. Update on methods and simulators for evaluation of noninvasive blood pressure monitors. J Clin Eng.. 1994;19:125-134. [Medline] [Order article via Infotrieve]

43. Mancia G, Parati G. Commentary on the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices: a critique of aspects related to 24-hour ambulatory blood pressure measurement. J Hypertens.. 1993;11:595-597. [Medline] [Order article via Infotrieve]

44. Reid J. Validation of blood pressure measuring systems. J Hypertens.. 1993;11:i-ii. [Medline] [Order article via Infotrieve]

45. White WB, Lund-Johansen P, McCabe EJ. Clinical evaluation of the Colin ABPM 630 at rest and during exercise: an ambulatory blood pressure monitor with gas-powered cuff inflation. J Hypertens.. 1989;7:477-483. [Medline] [Order article via Infotrieve]

46. White WB, Lund-Johansen P, McCabe EJ, Omvik P. Clinical evaluation of the Accutracker II ambulatory blood pressure monitor: assessment of performance in two countries and comparison with sphygmomanometry and intra-arterial blood pressure at rest and during exercise. J Hypertens.. 1989;7:967-975. [Medline] [Order article via Infotrieve]

47. White WB, Lund-Johansen P, Omvik P. Assessment of four ambulatory blood pressure monitors and measurements by clinicians versus intraarterial blood pressure at rest and during exercise. Am J Cardiol.. 1990;65:60-66. [Medline] [Order article via Infotrieve]

48. Northcote RJ, McKillop G, Wosormu D, Kesson E, Brown JJ. Accutracker: evaluation of a new ambulatory blood pressure monitor. Am J Noninvasive Cardiol.. 1990;4:276-281.

49. White WB, Pickering TC, Morganroth J, James GD, McCabe EJ, Moucha O, Hunter H. A multicenter evaluation of the A&D TM-2420 ambulatory blood pressure recorder. Am J Hypertens.. 1991;4:890-896. [Medline] [Order article via Infotrieve]

50. West JNW, Townsend JN, Davies P, Sheridan JJ, Stallard TJ, Davies MK, Littler WA. Effect of unrestricted activity on accuracy of ambulatory blood pressure measurement. Hypertension.. 1991;18:593-597. [Abstract/Free Full Text]

51. Barthelemy JC, Geyssant A, Auboyer C, Antonidas A, Berruyer J, Lacour JR. Accuracy of ambulatory blood pressure determination: a comparative study. Scand J Clin Lab Invest.. 1991;51:461-466. [Medline] [Order article via Infotrieve]

52. Hoegholm A, Edmak T, Kristensen KS, Bang LE, Madsen NH. Clinical evaluation of the Takeda Medical (A&D) TM 2420 ambulatory blood pressure monitor: practical experience and comparison with direct and indirect measurements. Scand J Clin Lab Invest.. 1992;52:261-268. [Medline] [Order article via Infotrieve]

53. Groppelli A, Omboni S, Parati G, Mancia G. Evaluation of noninvasive blood pressure monitoring devices SpaceLabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure. Hypertension.. 1992;20:227-232. [Abstract/Free Full Text]

54. Ng K-G, Small CF. Survey of automated noninvasive blood pressure monitors. J Clin Eng.. 1994;19:452-475. [Medline] [Order article via Infotrieve]

55. Ng K-G, Small CF. Survey of automated noninvasive blood pressure monitors. J Clin Eng.. 1995;20:469-479. Published comments and errata.

56. O’Brien E, Atkins N, Sheridan N, Killeen C, Costelloe R, Lydon S, O’Malley K. Evaluation of the Accutracker II non-invasive ambulatory blood pressure recorder according to the AAMI standard. J Ambul Monitor.. 1991;4:27-33.

57. Jyothinagaram SG, Watson D, Padfield PL. Suntech Accutracker ambulatory blood pressure monitor—clinical validation. J Ambul Monitor.. 1990;3:63-67.

58. Radaelli A, Coats AJS, Clark R, Bird R, Sleight P. The effects of posture and activity on the accuracy of ambulatory blood pressure recording: a validation of the Oxford Medilog system. J Ambul Monitor.. 1990;3:155-161.

59. Hope SL, Alun-Jones E, Sleight P. Validation of the accuracy of the Medilog ABP non-invasive blood-pressure monitor. J Ambul Monitor.. 1988;1:39-51.

60. Jamieson MJ, Fowler G, MacDonald TM, Webster J, Witte K, Lawson L, Crichton W, Jeffers AT, Petrie JC. Bench and ambulatory field evaluation of the A&D TM-2420 automated sphygmomanometer. J Hypertens.. 1990;8:599-605. [Medline] [Order article via Infotrieve]

61. Clark S, Fowlie S, Coats A, Radaelli A, van der Putt M, Bird R, Conway J. Ambulatory blood pressure monitoring: validation of the accuracy and reliability of the TM-2420 according to the AAMI recommendations. J Hum Hypertens.. 1991;5:77-82.

62. Clark S, Hofmeyr GJ, Coats AJS, Redman CWG. Ambulatory blood pressure monitoring during pregnancy: validation of the TM-2420 monitor. Obstet Gynecol.. 1991;77:152-155. [Medline] [Order article via Infotrieve]

63. Clark S, Fowler G, Pannarale G, Bebb G, Coats A. Age and blood pressure measurement: experience with the TM 2420 ambulatory blood pressure monitor and elderly people. Age Ageing.. 1992;21:398-403. [Abstract/Free Full Text]

64. O’Brien E, Mee F, Atkins N, O’Malley K. Inaccuracy of the Hawksley random zero sphygmomanometer. Lancet.. 1990;336:1465-1468. [Medline] [Order article via Infotrieve]

65. Conroy R, O’Brien E, O’Malley K, Atkins N. Measurement error in the Hawksley random zero sphygmomanometer: what damage has been done and what can we learn? BMJ.. 1993;306:1319-1322.

66. Conroy R, Atkins N, Mee F, O’Brien E, O’Malley K. Using Hawksley random zero sphygmomanometer as a gold standard may result in misleading conclusions. Blood Press.. 1994;3:283-286. [Medline] [Order article via Infotrieve]

67. Jamieson MJ, Petrie JC. Bench and ambulatory field evaluation of the A&D TM-2420 automated sphygmomanometer. J Hypertens.. 1991;9:578. Reply.

68. Conway J. Bench and ambulatory field evaluation of the A&D TM-2420 automated sphygmomanometer. J Hypertens.. 1991;9:577-578. [Medline] [Order article via Infotrieve]

69. O’Brien E, O’Malley K, Atkins N, Mee F. A review of validation procedures for blood pressure measuring devices. In: Waeber B, O’Brien E, O’Malley K, Brunner HR, eds. Ambulatory Blood Pressure. New York, NY: Raven Press; 1994:1-32.

70. Palatini P, Penzo M, Canali C, Pessina AC. Validation of the accuracy of the A&D TM-2420 Model 7 for ambulatory blood pressure monitoring and effect of microphone replacement on its performance. J Ambul Monitor.. 1991;4:281-288.

71. Imai Y, Sasaki S, Minami N, Munakakata M, Hashimoto J, Sakuma H, Sakuma M, Watanabe N, Imai K, Sekino H, Abe K. The accuracy and performance of the A&D TM 2421, a new ambulatory blood pressure monitoring device based on the cuff-oscillometric method and the Korotkoff sound technique. Am J Hypertens.. 1992;5:719-726. [Medline] [Order article via Infotrieve]

72. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Takeda TM-2420/TM-2020 determined by the British Hypertension Society protocol. J Hypertens. 1991;9(suppl 5):S17-S23.

73. O’Brien E, Atkins N, Mee F, O’Malley K. Evaluation of the SpaceLabs 90202 according to the AAMI standard and BHS criteria. J Hum Hypertens.. 1991;5:223-226. [Medline] [Order article via Infotrieve]

74. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the SpaceLabs 90207 determined by the British Hypertension Society protocol. J Hypertens. 1991;9(suppl 5):S25-S31.

75. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Novacor DIASYS 200 determined by the British Hypertension Society protocol. J Hypertens. 1991;9(suppl 5):S9-S15.

76. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Del Mar Avionics Pressurometer IV determined by the British Hypertension Society protocol. J Hypertens. 1991;9(suppl 5):S1-S7.

77. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the CH-Druck pressure ERKA ambulatory blood pressure measuring system determined by the British Hypertension Society protocol. J Hypertens. 1993;11(suppl 22):S1-S7.

78. O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Profilomat ambulatory blood pressure measuring system determined by the British Hypertension Society protocol. J Hypertens. 1993;11(suppl 22):S9-S15.

79. Mee F, Atkins N, O’Brien E. Validation of the Nissei DS-240 ambulatory blood pressure measuring system as determined by the British Hypertension Society protocol. J Hum Hypertens.. 1994;8:295. Abstract.

80. Taylor R, Chidley K, Goodwin J, Broeders M, Kirby B. Accutracker II (version 30/23) ambulatory blood pressure monitor: clinical validation using the British Hypertension Society and Association for the Advancement of Medical Instrumentation standards. J Hypertens.. 1993;11:1275-1282. [Medline] [Order article via Infotrieve]

81. White WB, Susser W, James G, Marra L, McCabe EJ, Pickering TG, Streeten DHP. Multicenter assessment of the QuietTrak ambulatory blood pressure recorder according to the 1992 AAMI guidelines. Am J Hypertens.. 1994;7:509-514. [Medline] [Order article via Infotrieve]

82. O’Shea JC, O’Neill A, O’Brien C, Murphy MB. Evaluation of the Tycos Quiet-Trak ambulatory pressure recorder using British Hypertension Society (BHS) protocol. Am J Hypertens.. 1994;7:118A. Abstract.

83. Shennan AH, Kissane J, De Swiet M. Validation of the SpaceLabs 90207 ambulatory blood pressure monitor in pregnancy. Br J Obstet Gynaecol.. 1993;100:904-908. [Medline] [Order article via Infotrieve]

84. O’Brien E, Mee F, Atkins N, Halligan A, O’Malley K. Accuracy of the SpaceLabs 90207 ambulatory blood pressure measuring system in normotensive pregnant women determined by the British Hypertension Society protocol. J Hypertens. 1993;11(suppl 5):869-873.

85. Cox J, Amery A, Clement D, Fagard R, Leonetti G, O’Brien E, O’Malley K, Mancia G, Parati G, Petrie J, Ritz E, Staessen J, Terozoli L, Webster J. Twenty-four hour blood pressure monitoring in the SystEur Trial. Ageing Clin Exp Res.. 1992;4:85-91.

86. Staessen J, Amery A, for the Advisory Board. APTH—a trial on ambulatory blood pressure monitoring and treatment of hypertension: objectives and protocol. Acta Cardiol.. 1993;48:25-42. [Medline] [Order article via Infotrieve]

87. Clement D, De Buyzere M, Duprez DD. Ambulatory blood pressure and prognosis: summary of ongoing studies. J Hypertens. 1991;9(suppl 8):S51-S53.

88. Imholz BPM, van Montfrans GA, Settels JJ, van der Hoeven GMA, Karemaker JM, Wieling W. Continuous non-invasive blood pressure monitoring: reliability of Finapres device during Valsalva manoeuvre. Cardiovasc Res.. 1988;22:390-397. [Medline] [Order article via Infotrieve]

89. O’Brien E, Atkins N. A comparison of the BHS and AAMI protocols for validating blood pressure measuring devices: can the two be reconciled? J Hypertens.. 1994;12:1089-1094. [Medline] [Order article via Infotrieve]

90. O’Brien E, Atkins N, Mee F, Coyle D, Syed S. A new audio-visual technique for recording blood pressure in research: the sphygmocorder. J Hypertens. In press.

91. Ludgate SM, Potter DC. European directives on medical devices. BMJ.. 1993;307:459-460.

92. O’Brien E. Accuracy and performance of the Terumo ES-H51, a new portable blood pressure monitor. Am J Hypertens.. 1994;7:118.

93. O’Brien E. Validity and reliability of 24h blood pressure monitoring in children and adolescents using a portable oscillometric device. J Hum Hypertens.. 1994;8:797-798. [Medline] [Order article via Infotrieve]

94. O’Brien E, Atkins N. Evaluation of the accuracy and reproducibility of the Takeda TM-2420 in the elderly. J Hum Hypertens.. 1995;9:205.

95. O’Brien E. Ambulatory blood pressure measurement and insulin resistance. Hypertension.. 1995;25:461.

96. O’Brien E. Automated blood pressure measurement in pregnancy. Am J Obstet Gynecol.. 1995;172:1328. [Medline] [Order article via Infotrieve]

97. O’Brien E, Atkins N. Blood pressure measurement using oscillometric finger cuffs. Anaesthesia. In press.




This article has been cited by other articles:


Home page
Diabetes CareHome page
T. B. Torbjornsdotter, G. A. Jaremko, and U. B. Berg
Nondipping and Its Relation to Glomerulopathy and Hyperfiltration in Adolescents With Type 1 Diabetes
Diabetes Care, February 1, 2004; 27(2): 510 - 516.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
F.M. McKevitt, A. Sivaguru, G.S. Venables, T.J. Cleveland, P.A. Gaines, J.D. Beard, and K.S. Channer
Effect of Treatment of Carotid Artery Stenosis on Blood Pressure: A Comparison of Hemodynamic Disturbances After Carotid Endarterectomy and Endovascular Treatment
Stroke, November 1, 2003; 34(11): 2576 - 2581.
[Abstract] [Full Text] [PDF]


Home page
CLIN PEDIATRHome page
A. M. Simckes, T. Srivastava, and U. S. Alon
Ambulatory Blood Pressure Monitoring in Children and Adolescents
Clinical Pediatrics, October 1, 2002; 41(8): 549 - 564.
[Abstract] [PDF]


Home page
CMAJHome page
J. E. Lewis, E. Boyle, L. Magharious, and M. G. Myers
Evaluation of a community-based automated blood pressure measuring device
Can. Med. Assoc. J., April 1, 2002; 166(9): 1145 - 1148.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
B. Frauchiger, H. P. Schmid, C. Roedel, P. Moosmann, and D. Staub
Comparison of Carotid Arterial Resistive Indices With Intima-Media Thickness as Sonographic Markers of Atherosclerosis
Stroke, April 1, 2001; 32(4): 836 - 841.
[Abstract] [Full Text] [PDF]


Home page
BMJHome page
E. O'Brien, B. Waeber, G. Parati, J. Staessen, and M. G Myers
Blood pressure measuring devices: recommendations of the European Society of Hypertension
BMJ, March 3, 2001; 322(7285): 531 - 536.
[Full Text]


Home page
Occup. Environ. Med.Home page
J. A Staessen, E. T O'Brien, L. Thijs, and R. H Fagard
Modern approaches to blood pressure measurement
Occup. Environ. Med., August 1, 2000; 57(8): 510 - 520.
[Abstract] [Full Text]


Home page
Nephrol Dial TransplantHome page
E. Lurbe and J. Redon
Assessing ambulatory blood pressure in renal diseases: facts and concerns
Nephrol. Dial. Transplant., November 1, 1999; 14(11): 2564 - 2568.
[Full Text] [PDF]


Home page
Arch Intern MedHome page
J. E. Naschitz, L. Loewenstein, E. Zuckerman, D. Yeshurun, J. A. Staessen, and E. O'Brien
Reliability of Self-Measured Blood Pressure for Research Purposes
Arch Intern Med, October 25, 1999; 159(19): 2365 - 2366.
[Full Text] [PDF]


Home page
HypertensionHome page
K. Seppa and P. Sillanaukee
Binge Drinking and Ambulatory Blood Pressure
Hypertension, January 1, 1999; 33(1): 79 - 82.
[Abstract] [Full Text] [PDF]


Home page
Arch Intern MedHome page
L. Thijs, J. A. Staessen, H. Celis, R. de Gaudemaris, Y. Imai, S. Julius, and R. Fagard
Reference Values for Self-recorded Blood Pressure: A Meta-analysis of Summary Data
Arch Intern Med, March 9, 1998; 158(5): 481 - 488.
[Abstract] [Full Text] [PDF]


Home page
HypertensionHome page
J. A. Staessen, L. Bieniaszewski, E. O'Brien, P. Gosse, H. Hayashi, Y. Imai, T. Kawasaki, K. Otsuka, P. Palatini, L. Thijs, et al.
Nocturnal Blood Pressure Fall on Ambulatory Monitoring in a Large International Database
Hypertension, January 1, 1997; 29(1): 30 - 39.
[Abstract] [Full Text]


Home page
BMJHome page
N. Prasad and C. Isles
Fortnightly Review: Ambulatory blood pressure monitoring: a guide for general practitioners
BMJ, December 14, 1996; 313(7071): 1535 - 1541.
[Abstract] [Full Text]


This Article
Right arrow Abstract Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by O’Brien, E.
Right arrow Articles by Staessen, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by O’Brien, E.
Right arrow Articles by Staessen, J.