Hypertension. 1996;27:1079-1089
(Hypertension. 1996;27:1079-1089.)
© 1996 American Heart Association, Inc.
Wave Propagation in Coupled Left VentricleArterial System
Implications for Aortic Pressure
David S. Berger;
Kimberly A. Robinson;
Sanjeev G. Shroff
From the Cardiology Section, Department of Medicine, University of
Chicago (Ill).
Correspondence to Sanjeev G. Shroff, PhD, University of Chicago Medical Center, Room M-507, MC-5084, 5841 S Maryland Ave, Chicago, IL 60637. E-mail sshroff@medicine.bsd.uchicago.edu.
 |
Abstract
|
|---|
Abstract The objective of this study was to examine the
effects
of wave propagation properties (global reflection coefficient,
G;
pulse wave velocity,
cph; and characteristic impedance,
Zo)
on the mechanical performance of the
coupled left ventriclearterial
system. Specifically,
we sought to quantify effects on aortic
pressure
(
Pao) and flow (
Qao)
while keeping constant other determinants
of
Pao
and
Qao (left ventricular
end-diastolic volume,
Ved,
and contractility,
heart rate, and peripheral resistance,
Rs).
Isolated rabbit hearts were subjected to
real-time, computer-controlled
physiological
loading. The arterial
circulation was modeled with a lossless
tube terminating in a complex
load. The loading system allowed
for precise and independent control of
all arterial properties
as evidenced by accurate
reproduction of desired input impedances
and computed left
ventricular volume changes. While propagation
phenomena
affected
Pao and
Qao
morphologies as expected, their
effects on absolute
Pao values were often contrary to the current
understanding.
Diastolic (
Pd) and
mean (
Pm)
Pao and stroke
volume decreased
monotonically with increases in
G,
cph,
or
Zo over wide ranges.
In contrast, these
increases had variable effects on peak systolic
Pao (
Ps): decreasing with
G, biphasic with
cph, and increasing with
Zo.
There was an interaction between
G and
cph such that
G effects
on
Pm and
Pd were augmented at higher
cph and vice versa. Despite
large changes in
system parameters, effects on
Pm and
Ps were
modest (<10% and <5%, respectively);
effects on
Pd were
always two to four times
greater. Similar results were obtained
when the single-tube model
of the arterial system was replaced
by an asymmetrical
T-tube configuration. Our data do not support
the prevailing hypothesis
that
Ps (and therefore ventricular
load)
can be selectively and significantly altered by manipulating
G,
cph,
and/or
Zo.
Key Words: pulse wave, propagation pulse wave, velocity blood pressure rabbit heart ventricular function compliance
 |
Introduction
|
|---|
The
arterial system
Zin, the
hydraulic load opposing ejection,
consists of two components: a steady
component composed of
Rs and a pulsatile
component consisting of distributed compliant
and inertial properties
(as well as, to a lesser extent,
Rs).
Together
with the cyclic nature of ventricular ejection, these
pulsatile
load properties give rise to wave propagation phenomena;
pressure
and flow waves originate at the left ventricle and propagate
along
the arterial tree at finite wave propagation
velocities, and
the waves are partially reflected at sites of impedance
mismatch
(eg, bifurcations, arterioles).
That wave reflections and propagation affect Pao
and Qao pulses has long been established, both
theoretically and experimentally; of this there is little
dispute.1 2 3 4 5 For example, the amplification of
Pao as it propagates along the aorta is one
manifestation of finite propagation velocities and
peripheral reflections.4 6 Also, the
morphological differences between Pao and
Qao are due to reflections, as evidenced by
their morphologies becoming more similar as reflections are
reduced.3 6 7 8 The shape of Pao
itself also contains some information regarding the magnitude of wave
reflections and their timing.9 It has been suggested that
in addition to the morphological changes in
Pao, reflected waves reaching the
ascending aorta in systole act to significantly augment measured
systolic pressure, ie, ventricular load, and that
this pressure elevation is greater with increased magnitude and/or
earlier return of reflections.10 11 12 13 14 15 On the basis of these
interpretations, it has been postulated that certain types of
hypertension, particularly those with elevated systolic
pressure, can be treated through reduction of reflections and wave
velocity.15 16 17 Although these hypotheses seem reasonable,
there exists no direct physical evidence that observed pressure
reductions are due to reduced reflections and/or reduced wave velocity
This is because of attendant experimental difficulties; studies
designed to address these issues, in both humans and intact
animals, leave uncontrolled one or more other factors that might
contribute to changes in pressure and flow. Moreover, recent
theoretical findings lead to the hypothesis that wave reflections
themselves do not necessarily increase the load faced by the
contracting left ventricle nor do they substantially affect mean
ventricular outflow, especially when compared with the
effects of Rs.8
In this study, we used an isolated heart preparation with real-time
loading of the left ventricle18 19 to examine
experimentally the effects of various wave propagation properties on
the mechanical performance of the coupled left
ventriclearterial system (ie,
Pao and Qao). This system
allowed independent control of arterial system load,
preload, and HR, thus permitting targeted changes in
arterial properties. The results presented in this
report will be compared with the previous model-based
findings8 and discussed in light of experimental work in
humans and intact animals. Finally, we will question the extent to
which reduction of wave reflections and/or cph
is beneficial to the left ventricle in terms of
Pao and SV.
 |
Methods
|
|---|
All protocols were reviewed and approved by the University of
Chicago
Institutional Animal Care and Use Committee and conform with
the
Guide for the Care and Use of Laboratory Animals
(National Institutes
of Health publication 85-23, revised 1985).
Experimental Preparation
Experiments were performed on hearts isolated from normal, adult
male rabbits (New Zealand White) weighing 2 to 3 kg. Rabbits were
preanesthetized with 5.0 mg/kg xylazine (Vedco) and 0.2
mg/kg atropine (Elkins-Sinn) and after 10 minutes were
anesthetized with 30 to 50 mg/kg ketamine (Kedalar,
Parke-Davis) and 1.0 mg/kg acepromazine (Vedco). Tracheotomy was
performed after anesthesia, and rabbits were artificially
ventilated (model 683, Harvard Apparatus) with room air at
a respiratory rate of 43 breaths/min and tidal volume of 25 to 30 mL.
After median sternotomy and ligation of great vessels, a metal cannula
connected to the perfusion system was inserted into the brachiocephalic
artery and immediately flushed with heparinized saline (3.0 mL, 1000
U/mL). Retrograde perfusion of the coronary arteries was then
begun at a constant perfusion pressure of 80 mm Hg and temperature of
37°C. The heart was perfused with oxygenated modified
Krebs-Henseleit solution,19 which was not recirculated.
Connective tissue was cut away and the heart removed from the chest
while being constantly perfused. Therefore, coronary
circulation was not interrupted at any time.
A thin latex balloon, secured at the end of a piston-cylinder
device, was positioned in the left ventricle via the mitral orifice. A
purse string tied around the mitral orifice secured the heart to the
piston-cylinder device attached to a linear motor. A
catheter-tip pressure transducer (model MPC-500, Millar
Instruments, Inc) was advanced into the left ventricle via a side port
in the piston-cylinder device. The piston position was sensed by a
linear voltage displacement transformer (model 294, Transtek). More
extensive details of the isolated heart setup can be found
elsewhere.18 19 All hearts were paced with unipolar
electrodes attached to the apex of the left ventricle.
Arterial System Model and Loading System
Fig 1A
depicts schematically the features of the
system that allow physiological loading of the
isolated left ventricle. Instantaneous Pv is
continuously monitored by the computer and acts as an input to the
model of arterial circulation, in this case a single,
uniform, elastic tube terminating in a complex load (Fig 1B
). The
arterial system tube parameters are
Zo, tube length (L), and
cph. The ratio
L/cph is
, the one-way
transmission time. Thus, propagation characteristics can be set by
prescribing either
or some combination of L and
cph. The relative contribution of reflected
waves at the left ventriclearterial system interface
is quantified by
G. It is important to
note that the model formulation is such that
Zo,
, and
G
can be varied independently. This allowed precisely controlled
experiments on wave propagation that are, for all practical purposes,
not possible in the real system because a change in a physical property
(eg, vessel wall stiffness) produces simultaneous changes
in Zo, cph,
and
G. The terminal load, a variation
on the three-element Windkessel,20 is given by
C, high-frequency resistance
(Ro), and Rs. The
tube and load are matched at high frequencies, therefore,

View larger version (22K):
[in this window]
[in a new window]
|
Figure 1. A, Schematic of real-time,
computer-controlled loading system. Command signal was updated and
data were sampled every 1.0 millisecond. B, Single-tube model with
complex terminal load. Abbreviations are as defined in the text.
|
|
 | (1) |
Following Campbell et al,21 this coupled system can
be described by the following delay-differential equation,
 | (2) |
 | (3) |
 | (4) |
where Pc is the pressure in the terminal
compliance chamber and constants
and ß are
given by
 | (5) |
With measured Pv as input,
Qao was calculated from numerical integration of
the system (fourth-order explicit Adams predictor-corrector
algorithm; step size=1.0 millisecond). The transitions from isovolumic
contraction to ejection and from ejection to isovolumic relaxation
occurred when Pv
Pao and
when Pv<Pao,
respectively. Calculated Qao was in turn
integrated to yield the desired reduction in instantaneous
Vv. The piston was moved exactly by an
amount equal to this desired volume change, and the whole process was
continued in real time with Pv samples taken
every 1.0 millisecond. After ejection, preset filling pressure
(Pf) and filling resistance
(Rf) were used for calculation of filling flow
(Qf):
 | (6) |
The onset of filling occurred when
Pf
Pv, and
filling continued until the desired Ved was
reached, at which time filling was stopped. Finally, HR was
independently controlled and systemic nervous and humoral influences
were not present. Thus, the experimental apparatus was
able to reproduce the conditions of the computer experiments described
in Berger et al8 with the same degree of precision and
independent control of arterial load
parameters.
Experimental Protocols
Several protocols were performed in which a single property of
the arterial system was altered, through adjustment of
specific model parameters, while everything else was kept
constant. These protocols were (1) protocol 
: adjusting
G(
) while keeping
Rs, Zo, and
cph constant; (2) protocol
cph: adjusting cph
while keeping |
G|,
Rs, and Zo
constant; and (3) protocol
Zo:
adjusting Zo while keeping
cph and Rs constant. A
total of 12 experiments were performed; six rabbits were used for
protocols 
and
cph combined and six for
protocol
Zo. Experimentally,
G(
) was altered by adjusting
only the value of C, thereby keeping the nonpulsatile
hydraulic load constant.8 Changes in
cph were accomplished by altering the wave
transmission time,
(Equations 2 through 4

), in equispaced
increments. These transmission times were converted to
cph changes assuming a fixed tube length of 0.1
m. Altering cph in this manner results in a
denser collection of points in the lower third of the
cph range.
In all protocols, HR was fixed at 120 beats/min.
Ved was first adjusted under isovolumic
conditions to yield end-diastolic
Pv of 5 to 10 mm Hg and thereafter was held
constant throughout the experiment. A single experimental "run"
consisted of eight different arterial loads in which only
one wave propagation property or parameter was changed, eg,
eight different
G(
) values for
protocol 
. The order of the eight predefined loads within a given
run was randomly chosen and administered by the computer. Data were
collected under numerical and physiological
steady-state conditions only. For the first of eight loading
conditions, the preparation was given 60 seconds to reach steady state;
the seven subsequent load changes were given at least 30 seconds. Once
steady state was reached, one ejection beat followed immediately by one
isovolumic beat from the same Ved were sampled.
This data collection scheme is illustrated in Fig 2
with
Pao and Pv (top),
Qao (middle), and Vv
(bottom) for the first two loading conditions in a run of protocol
cph. Data from several runs were collected
within each protocol, specifically, different
cph for protocol 
, different
G(
) for protocol
cph, and different
cph for protocol
Zo. The Table
contains
1, cph,
and Zo values of all runs for protocols 
,
cph, and
Zo, respectively.

View larger version (31K):
[in this window]
[in a new window]
|
Figure 2. Representative data record from
protocol cph showing
Pao (solid) and Pv
(dashed, top), Qao (middle), and
Vv (bottom) for the first two conditions from a
single run. After the start of the run, the preparation was allowed to
reach steady state with at least 60 seconds of undisturbed ejection. At
this time, one ejection beat followed immediately by one
Piso from the same end-diastolic
volume were sampled. After sampling, the load was changed by the
computer, and sampling took place after at least 30 seconds of
undisturbed ejection. This scenario was repeated for eight different
loads per run.
|
|
Occasionally, a premature ventricular contraction or other
arrhythmia would occur during ejection, thereby disturbing the
transition to steady state. Regardless of when this occurred, a full 30
seconds was given after the arrhythmia for steady state to be
reached. The total time for a single run was 5 to 6 minutes, short
enough so that natural degradation of the preparation was not a factor
within a given run. The total time required for all runs reported here
was approximately 1 hour, potentially long enough for degradation to
affect ventricular function. To minimize any effects of
degradation, we changed the order of protocols and runs within a
protocol from heart to heart.
Data Analysis
We subjected Pao and
Qao data to traditional wave reflection
analysis to obtain Zin(
)
and
G(
), where
Zin(
)=Pao(
)/Qao(
),
and
G(
)=[Zin(
)-Zo]/[Zin(
)+Zo].
The magnitude of the first and most significant
G(
) harmonic is denoted
1.
Zin(
)
=0
was compared with the desired Rs. If the
difference between the two values was greater than 2%, an indication
that either the preparation was not in steady-state ejection or the
closed-loop servocontrol was inadequate, that specific condition
was excluded from further analysis. This was a rare occurrence.
On the basis of Piso,
Pdev was determined as the difference between
peak and end-diastolic Piso.
Pdev was used to evaluate changes in left
ventricular contractility following each
loading condition and over time.
The effects of changes in wave propagation properties on
Pao and SV were analyzed. Specifically,
we were interested in Ps,
Pm, and Pd.
Data from different hearts were analyzed together on the basis
of percent changes from a common reference condition: smallest
1 (or equivalently largest C) for protocol

, largest cph for protocol
cph, and smallest
Zo for protocol
Zo.
 |
Results
|
|---|
First we evaluated the performance of the closed-loop
servocontolled
system. Fig 3A

shows model-derived
(analytic) magnitudes ||
Zin||
and
phases (
Zin) of
Zin(

) superimposed on values
calculated from
experimental measurements of protocol


. At all
frequencies,
an excellent agreement between model-derived and
experimental
Zin(

) can be seen.
Fig 3B

shows a close correspondence between
command
Vv (integral of
Qao)
imposed on the left ventricle and
actual
Vv
measured by the linear voltage displacement transformer
during
ejection.
Next we examined whether a variation in a specific wave propagation
property of the model yielded the desired variation in the
arterial system loading property.
Zin spectra and
G(
) magnitude
(|
G|) from one experiment are shown in
Figs 4
and 5
, respectively. In protocol

, diminished oscillations of both
|Zin| and
Zin (Fig 4
, left) as C increases indicate reduced wave reflections,
an observation confirmed by the data in Fig 5
(left). Furthermore,
unchanged high-frequency |Zin| and
little change in the zero-crossing of
Zin
(Fig 4
, left) confirm constant Zo and
cph. In protocol
cph, |Zin|
exhibited diminished oscillations as
cph increased and a more gradual convergence
toward a common Zo, whereas the
zero-crossing of
Zin was shifted toward
higher frequencies; ie, the system becomes more Windkessel-like
(Fig 4
, middle). Despite these changes in
Zin(
),
|
G| was completely unaffected by a change
in cph (Fig 5
, middle). In protocol
Zo, increasing
Zo yielded the expected increase in
|Zin| for all harmonics, with no change in
the zero-crossing of
Zin (Fig 4
, right).
Increasing Zo also yielded a reduction in
|
G| (Fig 5
, right), which is
consistent with the definition of
G(
).22 Also note
that Rs was unchanged for all conditions; thus,
Rs cannot be a factor in resultant
Pao and Qao changes.
Within any given run, Pdev was fairly constant,
never varying more than a few millimeters of mercury, and the time
courses of Piso were practically superimposable.
Over the course of the experiment (approximately 1 hour), a small but
noticeable (approximately 5%) decrease in Pdev
occurred. No other indications of diminished ventricular
performance such as altered coronary flow or elevated
end-diastolic Pv were observed in
the hearts from which these data are reported.
Protocol 

Fig 6
shows typical
Pv, Pao,
Qao, and Vv from a
single heart for control (middle), increased reflections (left), and
reduced reflections (right). Morphological changes in
Pao were as expected: increased
G(
) yielded increased pulse
pressure and end-ejection Pao
(Pee), with a systolic shoulder
developing at high
G(
).
Furthermore, Pao morphology begins to approach
that of Qao at low
G(
). Both
Ps and Pd increased with
decreasing
G(
), with relatively
larger changes in Pd.
Qao was also affected by
G(
). Increased
G(
) yielded a shorter ejection
period, higher peak flow, and narrower flow profile. Specifically,
ejection started and ended earlier with greater change in
end-ejection time. Fig 7A
shows
Ps, Pm, and
Pd with SV plotted against
1 over
the full range tested. The trends for this heart are clear: decreased
Ps and Pd with increasing
1 and changes in Pd becoming
larger at high
1; relatively constant
Pm, which eventually decreases at high
1; and changes in SV identical to those of
Pm, which is mandatory because
Rs and HR are unchanged.
The results from all six hearts in protocol 
are pooled in Fig 7B
, which shows percent changes (mean±SE) from the point of the
minimum reflection coefficient. Despite the different absolute pressure
levels between hearts, which were as great at 20 mm Hg, changes are
quite consistent along the entire range of increasing
1: a steady and small decrease in
Ps not exceeding 4%; a larger decrease in
Pd approaching 14%; and an almost constant
Pm until the highest values of
1, at which Pm decreased
less than 2%. Since the Pd decrement was much
greater than that of Ps, pulse pressure
increased by 44% over the range of increasing
1. Note
that although the choice of reference
1 is arbitrary,
the directional changes observed are not dependent on the reference
value of
1. This comment applies to all subsequent
percent change results.
Protocol
cph
Fig 8
depicts typical
Pv, Pao,
Qao, and Vv from a
single experiment for cph=7.58 m/s (left),
cph=3.79 m/s (middle), and
cph=2.16 m/s (right). Altering
cph, which affects the timing of the
reflected wave, has clear, consistent effects on
Pao morphology: distinct diastolic
oscillations with reduced cph and a
gradual elimination of these oscillations as the reflected
wave arrives earlier with increasing cph. Both
peak Qao and ejection period were affected
little by cph, whereas the flow profile
was more symmetrical at low cph. Fig 9A
shows Ps,
Pm, Pd, and
SV over the full range of cph. Both
Pm and Pd showed initial
rapid decreases with increasing cph (SV followed
Pm), leveling off at high velocities.
Ps showed a biphasic response, with a minimum
value occurring at approximately 5 m/s.

View larger version (28K):
[in this window]
[in a new window]
|
Figure 8. Results from one experiment in protocol
cph: Pao and
Pv (top), Qao (middle),
and Vv (bottom) for
cph=7.58 m/s (left),
cph=3.79 (middle), and
cph=2.16 (right).
|
|
Percent changes (mean±SE) in Ps,
Pm, and Pd for all
six experiments in protocol
cph are shown in
Fig 9B
, with the highest velocity as the reference condition. These
changes are remarkably consistent, with
Pd being the most sensitive, showing a 12%
change across the entire cph range. Changes in
Ps and Pm were smaller:
less than 4% for the former and 5% for the latter. Consequently,
pulse pressure consistently increased with increasing
cph. The biphasic Ps
response was evident in all hearts.
Combined 
and
cph
Percent changes in Ps,
Pm, and Pd due to
changes in
1 at three different
cph and due to changes in
cph at three different
1 are
shown in Fig 10
. For each experiment, a single
reference condition (lowest
1 and highest
cph) was used for computation of these percent
changes. Pd and Pm
decrements after an increase in
1 were augmented as
cph increased (Fig 10A
). Similarly
Pm and Pd decrements in
response to increasing cph were augmented by
1 increases (Fig 10B
). Regarding
Ps, its biphasic response to changing
cph also was augmented, and the point of minimum
Ps was shifted to lower
cph by increased
1 (Fig 10B
). In
contrast, different cph had no effect on the
slope of the Ps-
1 relationship
until large values of
1 were reached. Also note that for
any combination of
1 and cph
changes, the percent change in Pd was by far the
largest of the three pressures.
Protocol
Zo
Fig 11
shows typical
Pv, Pao,
Qao, and Vv for a
single experiment for Zo=5.0 mm Hg·s/mL
(left), Zo=7.0 mm Hg·s/mL (middle), and
Zo=10.0 mm Hg·s/mL (right). Peak
Qao decreased while ejection period increased
(combined earlier start- and later end-ejection times) with
increasing Zo. Increasing
Zo over a wide range yielded increased pulse
pressure through combined decreases in Pd and
increases in Ps (Fig 12A
). The
pooled data from six hearts in Fig 12B
again show very
consistent percent changes for all pressures. As in protocols

and
cph,
Pd was the most sensitive variable, changing
almost 16% over the Zo range. The percent
change in Ps was also similar in magnitude to
those from protocols 
and
cph,
increasing 4% as Zo increased over the range.
The percent change in Pm (or equivalent in SV)
was large compared with those from protocols 
and
cph, decreasing 10% over the
range.

View larger version (28K):
[in this window]
[in a new window]
|
Figure 11. Results from one experiment in protocol
Zo: Pao
and Pv (top), Qao
(middle), and Vv (bottom) for
Zo=5.0 mm Hg·s/mL (left),
Zo=7.0 mm Hg · s/mL (middle), and
Zo=10.0 mm Hg · s/mL (right).
|
|
Unlike in protocols 
and
cph, in
which rates of pressure changes varied over their respective ranges
(ie, biphasic or eventually plateauing), the changes in pressures due
to changes in Zo were linear over the entire
range. Changes in cph shifted the
Pd-Zo and
Pm-Zo relationships in a
parallel manner (Fig 13
); the effect on the
Ps-Zo relationship was
more subtle because of the biphasic response of
Ps to changes in cph.

View larger version (20K):
[in this window]
[in a new window]
|
Figure 13. Percent changes (mean±SE) in
Ps (top), Pm (middle),
and Pd (bottom) from a single reference
condition for three different values of cph in
protocol Zo.
|
|
 |
Discussion
|
|---|
Our overall goal is to understand better the ways in which wave
reflections
and
cph affect
cardiovascular performance, from both
mechanical
and energetic perspectives. Here we report the effects of
G(

),
cph,
and
Zo on
Pao and
Qao generated in
the coupled left ventriclearterial
system. This is
the first experimental study that examines systematically
the
independent effects of the three aforementioned wave propagation
properties
under rigorously controlled conditions.
Adequacy of Experimental Control
For study of the direct effects of wave propagation phenomena on
Pao and Qao,
several experimental constraints must be met8 : (1)
complete control of arterial system properties such that
G(
) and
cph can be altered independently of each other
and Zo; (2) ability to alter propagation
properties without affecting the steady component of the
arterial load, ie, constant
Rs; (3) control of preload,
Ved, and HR; (4) elimination of
neurohumoral feedback; and (5) a heart whose contractile properties do
not change over time.
The isolated heart preparation readily fulfilled requirements 3 and 4.
Data presented in Figs 3 through 5

indicate that requirements 1
and 2 are also met. Regarding requirement 5, a genuine potential source
of error was unavoidable degradation of the preparation, some of which
was evident in the decreased Pdev as experiments
progressed. Although the magnitude of this decrease was small (
5%),
it may have affected late in an experiment the sensitivities of
Pao and Qao to load
changes. To minimize the systematic effects of such degradation, we
rearranged the order of protocols and runs within a protocol between
hearts.
Are Results Specific to the Chosen Arterial
Model?
The single-tube model treats all peripheral
reflections as arising from one equivalent reflection site. To test the
possibility that multiple reflection sights will yield different
results, we loaded two hearts with the asymmetrical T-tube
model21 23 consisting of two parallel circulations:
head-end and body-end. Protocols 
,
cph, and
Zo were implemented by changing
parameters of the two circulations
simultaneously and in proportion. Results with T-tube
loading (Fig 14
) were the same as those with
single-tube loading. Thus, with uniform changes in the
arterial system properties, as would occur in most cases,
the number of reflection sites does not seem to matter. The issue of
whether the same results will emerge with nonuniform changes (eg,
localized alterations resulting in new, discrete reflection sites)
needs further investigation.
Extending this argument further, we believe that results would be
minimally affected if more complex models of the arterial
system were used, eg, more branches, elastic and/or geometric taper,
lossy tubes, etc. It is sufficient to use a minimal model that
reproduces Zin accurately under a variety of
vasoactive conditions and allows for independent changes in reflection
coefficient and wave velocity. Single- and T-tube models with uniform,
lossless tubes and complex terminal load satisfy these
criteria.22 24 25 26
Reflection Coefficient and Pulse Wave Velocity: Comparison With
Model-Based Study
Except for Pd, which was by far the
most sensitive variable, changes in all pressures and SV were
modest, especially considering the wide ranges of
1 and
cph examined. These results, along with the
morphological changes in Pao and
Qao due to changes in
G(
) and
cph, are entirely consistent with
those found in the model-based study.8 The physical
reasons for these changes were discussed in detail
elsewhere.8 Briefly, alterations in
G(
) and
cph result in a redistribution of pressure
between diastole and systole such that
Pm/SV is maintained (a consequence of
constant HR and Rs). So, with a reduction in
1, for example, the simultaneous
reduction of diastolic pressure decay and slight increase
in SV yield a proportional increase in Pm and a
redistribution of pressure from diastole to systole such
that Ps increases slightly.
Characteristic Impedance
Unlike protocols 
and
cph,
changes in Zo simultaneously alter a
second arterial system property, namely
G(
). Thus, the effects of
Zo on pressure and flow can be considered the
result of changes in both Zo and
G(
), and it is instructive to
compare the results in Fig 12B
with those in Fig 7B
. The increase in
Zo from 3.0 to 10.0 mm Hg·s/mL yielded a
decrease in
1 from 0.62 to 0.20 (Fig 5
). Over this range
of decreasing
1, Ps
increased by approximately 3%, Pm was
unchanged, and Pd increased by approximately 6%
(Fig 7B
). With protocol
Zo,
corresponding changes after an increase in Zo
were (Fig 12B
) approximately 4% increase in
Ps, approximately 9% decrease in
Pm, and approximately 15% decrease in
Pd. Thus, changes in
1
certainly contribute to Ps changes in protocol
Zo. Since
Pm was unaffected by
1 over this
range, the Pm changes in protocol
Zo are likely due to changes in
Zo alone. Finally, note that the directional
changes in Pd with decreasing
1
are opposite between protocols 
and
Zo. Therefore, the effects
of Zo alone (ie, with invariant
G) on Pd are expected
to be greater than these observed in protocol
Zo (
Zo and
1).
Interaction Among
1,
cph, and
Zo Effects
Results presented in Fig 10
indicate that the
sensitivities, and in some cases the directionality, of pressure
responses due to changes in one parameter are greatly
modified by changes in another. In general, effects of reflections are
augmented at higher wave velocities and vice versa. On the other hand,
percent changes in Ps,
Pm, and Pd due to
changes in Zo are unaffected by
cph (ie, no slope change in Fig 13
). In contrast
to the experiments described here,
(patho)physiological events in the intact
circulation (eg, exercise, disease processes, pharmacological
treatment, aging) can simultaneously affect multiple
arterial system properties. That is, certain
arterial system properties tend to correlate, because of
either shared underlying physical properties or dependence on common
physiological responses. For example, aortic
cph and Zo tend to follow
one another because both are inversely related to vessel wall
stiffness.1 By traversing along and between the families
of curves (Figs 10
and 13
), changes in pressures due to
simultaneous changes in wave propagation properties can be
predicted. In doing so, one would still observe that changes in
Ps are much smaller than those in
Pd.
Comparison With Previous Studies
Morphological changes in Pao and
Qao observed in this study are
consistent with those of previous reports. Specifically,
development of a systolic shoulder in
Pao is a classic morphological indicator of
increased contribution of reflected waves,4 and the
disappearance of diastolic oscillations and
gradual similarity between Pao and
Qao morphologies with reduced reflections has
been observed in human and intact animal
experiments.6 7
With respect to pressure and flow magnitudes, the general consensus is
that wave reflections act to increase the pressure load on the
ventricle. Consequently, reducing the magnitude of the reflections,
delaying the return of reflections, or both have been associated with
significantly reduced Ps with relatively little
effect on Pd.11 12 13 Our
results do not agree with these observations. We show small changes in
Ps and Pm (or SV), and in
some cases increases, with large reductions in
G(
) and
cph. Furthermore, changes in
Pd are much larger than changes in both
Ps and Pm.
The role of wave propagation has been investigated with one of two
approaches: intact animal experiments or modeling experiments in which
both the arterial system and left ventricle are
represented by mathematical constructs. The present
study represents a hybrid of the two: a real heart coupled to a
physical/mathematical model of the arterial system. We will
reconcile our observations with those in the literature on the basis of
two issues: (1) the left ventricle is neither an ideal flow nor an
ideal pressure source27 and (2) properties other than wave
reflection and propagation are often left uncontrolled.
Regarding the first issue, making any change in the
arterial system, however small, will change
Zin(
) in some way, thereby
changing the hydraulic load into which the heart ejects. Therefore, the
effects on Pao and Qao
arising from changes in
G(
),
cph, and/or Zo are
twofold: ejection and subsequent wave modification. In the dynamically
coupled system, model or animal, the heart interacts with the load
until steady-state solutions of Pao and
Qao emerge. Therefore, all aspects of
Pao and Qao are subject
to change when the load changes, as seen in Figs 6
, 8
, and 11
. Changes
in
G(
) and
cph provide information regarding only the
nature of the backward wave relative to the forward wave; effects on
the forward wave itself are not addressed. Thus, one cannot simply make
predictions a priori about the effects of wave propagation properties
on measured Pao and Qao
based solely on the known effects of these properties on the
relationship between forward and backward waves; the effects on
ejection should not be overlooked.
Theoretical Model Studies
Although many studies have used wave transmissiontype
arterial models, only those in which a dynamic
ventricular source is coupled to the arterial
model are suitable for comparison with the results presented
here. McIlroy and Targett28 showed both elevated
Ps and Pd with increased
|
G|. Since
|
G| was increased in their study by
increasing Rs, both wave reflections and
the steady component of load contributed to the observed changes in
pressure.8 Using a multibranching arterial
model, Fitchett29 decreased arterial
compliance by increasing the elastic modulus for all segments (which
corresponds to a simultaneous increase in
cph and Zo and possibly a
decrease in
1) and observed pressure changes that are
consistent with our data (Figs 10
and 13
). However, his
observation that Ps changed more than
Pd after increased wave reflections is contrary
to our results. Although the difference cannot yet be fully explained,
it may be due to the way in which wave reflections were manipulated:
80% occlusion and a fourfold increase in elastic modulus of a single
segment. As mentioned before, such nonuniform, discrete changes in wave
reflections will be the subject of future investigations. Finally,
Latson et al30 and Burkhoff et al31
analyzed Pao generated by a left
ventricle model connected to measured Zin and
Pao generated by the same model left ventricle
connected to an equivalent three-element Windkessel. Their
Windkessel equivalent corresponded to not only an increase in
cph to infinity but also a reduction in
|
G|, a consequence of maintaining the
total arterial compliance (ie, redistribution from tube to
load). Results from these studies are consistent with our data,
keeping in mind that these changes in cph and
|
G| have competing effects on
Pd and Pm and
variable effects on Ps (Fig 10
).
Isolated Heart Studies
Most isolated heart studies have used the three-element
Windkessel model, in which cph is
infinite.27 32 33 34 Nevertheless, changes in Windkessel
compliance, Cw, are equivalent to changes
in
G(
); decreased
Cw results in increased
|
G|. In all these studies, a reduction in
Cw caused the following: a small decrease in
Pm and SV, a decrease in
Pd, and either an
increase27 34 or a biphasic response32 33 in
Ps. Changes in Ps
relative to Pd were always smaller. These
observations are entirely consistent with our results from
protocol 
at the highest cph (15.15 m/s,
Fig 10A
).
In contrast to the Windkessel loading, Kirkpatrick et al18
successfully loaded the isolated heart with wave transmission models;
our loading system is identical to theirs. Although Kirkpatrick et al
primarily focused on presenting a new methodology, one can observe
effects of load compliance (Fig 7
in Kirkpatrick et al) and wave
velocity (Fig 6
in Kirkpatrick et al) on Pao and
SV that are similar to what we report here.
Intact Heart Studies
Several methods have been used to alter arterial wave
propagation properties in the intact circulation: providing an
alternative ejection path via either an artificial
Windkessel35 or a stiff tube placed in the ascending
aorta36 ; applying plaster or Lucite ferrules to various
portions of the aorta and large arteries37 ; and replacing
a section of native aorta with glass38 or
Tygon39 tubing. These manipulations cause an increase in
"aortic" wall stiffness, resulting in an increase in
cph and in some cases an increase in
Zo. All of these studies show that
increasing aortic stiffness in these manners yields decreased
Pd, increased pulse pressure, and a
relatively smaller reduction in SV, observations consistent
with the results of the present study. However, increments in
Ps were significantly greater than those from
the present study, their magnitudes being more comparable to the
decrements in Pd. These magnitude
discrepancies can be explained by changes in factors other than those
associated with wave propagation. For example, Randall et
al36 showed trends of increasing Rs
and Ved (inferred from their left
ventricular end-diastolic diameter and
pressure) with increased aortic stiffness. Similarly, it can be
inferred from the data of Kelly et al39 that
Rs and Ved increased by
approximately 20% and 13%, respectively. Since
Pao is highly sensitive to
Rs and Ved, even
small increases in these quantities can significantly augment
Ps, thus reconciling the
discrepancies.
Another class of studies uses drug effects, aging, and disease
processes to make inferences regarding the effects of wave propagation
phenomena. In general, elevated systolic pressure (eg, isolated
systolic hypertension in the elderly) is accompanied by
increases in wave reflections and/or increased
cph.40 41 42 Conversely, vasodilator
drugs, such as nitroprusside and nitroglycerin, that
lower systolic pressure also tend to reduce wave
reflections.10 11 40 43 On the basis of these
observations, it has been hypothesized that reducing wave reflections,
especially with greatly elevated Ps, is
beneficial in that Ps can be reduced
significantly with little change in Pd.
Our results are clearly inconsistent with these hypotheses.
Once again, we propose that changes in factors other than wave
reflections are responsible for this discrepancy. For example, Yaginuma
et al11 report that in normotensive patients without
cardiac disease, nitroglycerin reduces
Ps, Pm, and
SV significantly without any change in
Pd. They attribute these changes to a
reduction in peripheral reflections only; since
Rs, Zo,
cph, Cw,
and HR did not change. We offer an alternative explanation. All
pressure and SV changes observed by Yaginuma et al can be reproduced
simply by a 20-mL reduction of Ved,
consistent with vasodilator-induced venous pooling, without
invoking any other cardiovascular changes. For a normal
human left ventricle, this change in Ved would
cause left ventricular end-diastolic
pressure to decrease by less than 4 mm Hg,44 a pressure
change that was actually observed by Yaginuma et al. Similar
concerns are applicable to subsequent studies regarding effects of
vasodilator-mediated changes in wave reflections and consequent
alterations in Pao.12 13 14 16 That
wave reflections are reduced with nitroglycerin and
other vasodilators is not disputed; however, this reduction of
reflections in itself cannot preferentially reduce
Ps.
Physiological Significance
One must keep three things in mind when extrapolating to the
intact circulation our results regarding the effects of wave
propagation properties on Pao and
Qao. First, we purposely examined very wide
ranges of
1, cph,
and Zo. Typical changes in these
parameters under various
(patho)physiological conditions are smaller,
especially for cph (as much as threefold change)
and Zo (as much as twofold change). Therefore,
in the intact circulation, expected changes in
Pao and Qao would be
smaller than the extremes shown in this study. Second, effects are
small for Pm (or SV) and
Ps. Furthermore, since
1, cph, and
Zo usually change in the same direction and
their individual effects on Ps compete, the net
response of Ps may be attenuated when two or
more properties change simultaneously. Third,
Pd is by far the most affected, especially with
combined changes in which individual effects are in the same direction.
Thus, our data do not support the hypothesis that
Ps (hence, ventricular load) can be
selectively and significantly reduced by reducing reflections.
In this study, we evaluated the relevance of wave propagation and
reflection in normal hearts and solely in terms of the effects on
Ps, Pm,
Pd, and SV. It is worthwhile to also
examine the effects of wave reflections on other variables, eg,
left ventricular relaxation and filling, coronary
flow, and myocardial oxygen consumption. Energetics and
coronary flow are particularly interesting because of the
greater sensitivity of Pd. Finally, the
effects of wave reflections in diseased and hypertrophied hearts as
well as the role of wave propagation phenomena, if any, in the
remodeling process also remain to be elucidated.
 |
Selected Abbreviations and Acronyms
|
|---|
| C |
= |
terminal load compliance of arterial system |
| cph |
= |
pulse wave velocity |
G |
= |
global reflection coefficient |
1 |
= |
magnitude of the first and most significant
G( ) harmonic |
| HR |
= |
heart rate |
| Pao |
= |
aortic pressure |
| Pd |
= |
minimum diastolic aortic pressure |
| Pdev |
= |
peak developed left ventricular pressure in isovolumic beat |
| Piso |
= |
left ventricular pressure in isovolumic beat |
| Pm |
= |
mean aortic pressure |
| Ps |
= |
peak systolic aortic pressure |
| Pv |
= |
left ventricular pressure in ejecting beat |
| Qao |
= |
aortic flow (left ventricular outflow) |
| Rs |
= |
peripheral resistance |
| SV |
= |
stroke volume |
| Ved |
= |
left ventricular end-diastolic volume |
| Vv |
= |
left ventricular volume |
 |
= |
angular frequency |
| Zin |
= |
arterial system input impedance |
| |