(Hypertension. 1997;29:1119-1125.)
© 1997 American Heart Association, Inc.
Articles |
From the Departments of Thoracic and Cardiovascular Surgery (S.Y.K.) and Medicine (D.E.E.), Loyola University Medical Center, Maywood, Ill.
| Abstract |
|---|
|
|
|---|
Key Words: baroreflex posture complex demodulation
| Introduction |
|---|
|
|
|---|
In addition to power spectral analysis, oscillations in RR interval and BP can also be assessed by complex demodulation (CDM). Whereas power spectral analysis establishes the presence of discrete frequency components within a time series, CDM assumes that periodic oscillations exist within a given frequency range and quantifies the amplitude of these oscillations as a function of time.15 Whereas power spectral analysis assumes that the amplitudes of oscillations are stationary over time, CDM provides a continuous assessment of the amplitudes of oscillations even when their amplitudes vary over time.15 Several recent studies have shown that CDM can be used for assessing variability in BP and RR interval.16 17 18 Furthermore, complex-demodulated amplitudes of BP and RR interval of the low-frequency components (0.04 to 0.14 Hz) oscillated in phase with a significant cross-correlation.17 Therefore, our purpose in this study was to determine whether CDM could also be used to provide a dynamic assessment of baroreflex sensitivity. We performed studies using both simulated and actual data obtained from healthy volunteers.
| Methods |
|---|
|
|
|---|
Real-Time Data Acquisition and Analysis
All subjects were instructed to avoid beverages containing
alcohol or caffeine for 24 hours preceding the study. At least 2 hours
after a light meal, subjects were studied in an air-conditioned
(23°C) and light-attenuated laboratory between 2 and 4
PM. A lead II electrocardiogram and
noninvasive beat-to-beat BP from the right middle finger (Finapres
2300, Ohmeda) were continuously monitored. The analog signals were
digitized at a rate of 250 Hz (Dataq) and stored on a 486 personal
computer (Everex) for off-line analysis. After a 10-minute
acclimation period with subjects in the supine position,
recordings were made for 10 minutes with subjects in the supine
position and for 12 minutes in the standing position. During both
supine and standing positions, arterial BP was monitored
with the cuffed finger at heart level to eliminate any hydrostatic
gradients. The position of the finger was marked with a skin electrode
at the fourth intercostal space in midaxillar line. The BP values of
the missing beats (2 to 3 beats per every 50 to 70 beats) caused by
automatic calibration of the Finapres were estimated by linear
interpolation. Commercially available software (Dataq) was used for
determination of the RR interval and peak systolic BP during
each cardiac cycle. Markers were placed in the electrocardiographic and
BP waveforms indicating the positions of the detected R wave and
systolic peaks, and any errors in the peak detection were
edited manually.
Power Spectral Analysis
A fast Fourier transform was applied to a series of 512
consecutive and stationary cardiac intervals and the systolic
arterial BP that corresponded to each cardiac
interval.19 The first 2 minutes after standing was
excluded from the analysis. Stationarity was defined as a
difference of less than 5% in the spectral components calculated in
two successive 256-beat series.8 The fast Fourier
transform was used to divide the overall variability of BP and cardiac
interval into frequency components. The low-frequency power was
calculated from integrating the curve from 0.04 to 0.14 Hz.
Cross-spectral analysis not only gives variability as a
function of frequency but also quantifies covariation between
systolic BP and cardiac interval in terms of modulus (gain),
coherence (the amount of linear coupling between the two
variables), and phase (time shifts) in a specific frequency band.
Baroreflex sensitivity was assessed by the cross-spectral modulus
(milliseconds per millimeter of mercury) of the transfer function
between variations in the BP and cardiac interval. The arithmetic means
of the moduli with a coherence value greater than 0.5 were used for
calculation of baroreflex sensitivity.7 8 9 10 11 12
Sequential Analysis of Spontaneous Variations in BP and
RR Interval
Beat-to-beat changes in systolic BP and RR interval were
computed over 512 cardiac cycles using the difference equations. The
computer algorithm then identified all sequences of three or more
successive cardiac beats in which there were concordant increases or
decreases in systolic BP by at least 1 mm Hg per beat and
RR interval by at least 4 milliseconds per beat. A linear regression
between the systolic BP values and the following RR intervals
(ie, a one-beat delay) was applied to each of the sequences. When the
regression analysis yielded a correlation coefficient higher
than .9, the slope (milliseconds per millimeter of mercury) was assumed
to reflect baroreflex sensitivity. The average regression slope for all
sequences was taken as the baroreflex sensitivity for the entire data
sampling period.9 13 The algorithm found 15±8 sequences
that fit the above criteria in the supine position and 23±11 in the
standing position.
Complex Demodulation
CDM is a time-local version of harmonic analysis that
provides time-dependent changes in amplitude and phase of a particular
frequency component as a function of time (see
Appendix).15 Briefly, the process of CDM involves shifting
the frequency band of interest to zero by multiplying the original
signal with a complex sinusoid at the center frequency of the spectral
region of interest (fo). The resultant complex signal is
then low-pass filtered and converted to a polar form to produce
amplitude and phase, as a function of time, of the component at
fo. The amplitude and phase variation of CDM indicate the
intensity of the signal around and the relative frequency deviation
from fo, respectively.15 Because of this
characteristic, CDM can be used for examination of dynamic changes in
the amplitude of oscillations in heart rate and
BP.17
In the present study, the time-dependent changes in the amplitude of the low-frequency components of the RR interval and systolic BP were assessed by CDM. A center frequency (fo) of 0.09 with low-pass filter corner frequency of 0.05 Hz (a 12-pole Butterworth) was applied to 512 cardiac cycles to produce a frequency range of 0.04 to 0.14 Hz. A Butterworth filter was used for demodulation because this filter has a maximally flat response curve. However, the filter causes phase shifts, which lead to a delay in the time domain, and transient responses (ringing) at the beginning of the filtered data series.16 To reduce these artifacts and eliminate the phase shift, we padded the complex-demodulated data series on each end with their mirror images and passed both backwards and forwards through the filter.19 In the event that there was no variability component within 0.4 to 0.14 Hz, CDM would yield unreliable (near zero) amplitude values.18 To minimize the contribution of any unreliable amplitude values to the determination of baroreflex sensitivity, we applied a five-point moving average filter to the complex-demodulated data.19 After both complex-demodulated data sets were filtered, the ratio of the instantaneous amplitude of the RR interval to the amplitude of systolic BP was calculated and defined as baroreflex sensitivity (milliseconds per millimeter of mercury).
Comparison of Baroreflex Sensitivity Values by Different
Methods
Since the validity and reliability of CDM when applied to the
assessment of baroreflex sensitivity in the time domain are unknown, we
first tested its validity with simulated data and then evaluated its
reliability with data from the human subjects. To test the validity of
CDM, we simulated beat-to-beat systolic BP and RR interval
representing supine and standing periods with a set of
cosine functions. CDM of the simulated data was performed with the same
set of parameters (center frequency, low-pass filter corner
frequency, and frequency bandwidth) as described above. To examine the
reliability of CDM-derived baroreflex sensitivity, we compared the
values obtained with this method with the corresponding values obtained
from power spectral analysis and sequential analysis.
The quantitative relationship between the values obtained with CDM and
power spectral analysis or CDM and sequential analysis
was evaluated by linear regression analysis. A one-way ANOVA
for repeated measures was used for comparison of the baroreflex
sensitivity values calculated by the three methods. For each method of
analysis, a paired t test was used for comparison of
the baroreflex sensitivity between the two postures. To investigate the
temporal relationships between the amplitude of CDM-derived
oscillations in BP and the amplitude of CDM-derived
oscillations in RR interval, we performed a
cross-correlation analysis and expressed the results as the
coefficient at zero lag.17 20 All results are
presented as mean±SD; significance was considered to be
present at a value of P<.05.
| Results |
|---|
|
|
|---|
x0.09t)+900 milliseconds
and SBPsupine=5 cos(2
x0.09t)+120 mm Hg;
and RRstanding=50 cos(2
x0.09t)+700 milliseconds and
SBPstanding=8 cos(2
x0.09t)+140 mm Hg. The results
of baroreflex sensitivity values calculated from CDM and power spectral
analysis of the simulated data are shown in Fig 1C
|
To verify the capability of CDM to detect dynamic changes in amplitude and frequency, we performed time and frequency resolution tests.17 18 20 The time resolution was determined by comparing baroreflex sensitivity values obtained from CDM using short segments (1 to 30 seconds) of the simulated data. The baroreflex sensitivity reached 98% of its steady-state value within 20 seconds. The frequency resolution was tested with simulated data that had a constant amplitude (100) and a linearly increasing frequency from 0 to 0.5 Hz over a period of 1000 seconds. CDM gave an amplitude of 100 only when the instantaneous frequency was between 0.0525 and 0.127 Hz. The amplitude decreased by 95% at a frequency higher than 0.032 Hz (transitional bandwidth) apart from the upper and lower limits of the frequency band.
Postural Studies in Human Subjects
Fig 2
shows a time series plot of RR interval (Fig 2A
) and systolic BP (Fig 2B
) for a
representative subject. A change in posture from supine
to standing increased heart rate from 62 to 78 beats per minute and
systolic BP from 130 to 141 mm Hg. Fig 2
also shows
CDM-derived baroreflex sensitivity values in the time domain. In
addition, baroreflex sensitivity values determined by power spectral
analysis are shown as a function of frequency for the supine
(Fig 2D
) and standing (Fig 2E
) positions. The values of CDM-derived
baroreflex sensitivity in this subject averaged over the
sampling period were 17.4 (supine) and 7.9 (standing) ms/mm Hg and
were similar to those obtained by power spectral analysis (17.0
and 7.9 ms/mm Hg, respectively) and sequential analysis (18.0
and 7.3 ms/mm Hg). The reliability of CDM-derived baroreflex
sensitivity measurements in human subjects was examined by comparing
the results of CDM with the results of power spectral analysis
and sequential analysis (Table
). With subjects
in the supine position, CDM-derived baroreflex sensitivity (averaged
over 512 beats) was 13.9±5.2 ms/mm Hg and not significantly different
from the baroreflex sensitivity calculated by power spectral
analysis (13.7±6.7) or sequential analysis
(14.3±6.5). Likewise, with subjects in the standing position, the
CDM-derived baroreflex sensitivity was 7.3±2.8 ms/mm Hg and not
significantly different from the value calculated by power spectral
analysis (7.0±3.0) or sequential analysis
(7.2±2.8).
|
|
A scattergram of the baroreflex sensitivity values derived from CDM and
power spectral analysis is depicted in Fig 3
.
The results of linear regression analysis showed the two
methods to be highly correlated (r=.97, P=.0001).
A significant correlation was also found between the CDM-derived
baroreflex sensitivity and the corresponding values determined by
sequential analysis (r=.98, P=.0001). The
cross-correlation analyses showed that variations of the
amplitudes of RR interval and systolic BP
oscillations derived by CDM occurred with a significant
temporal relationship in all subjects. The cross-correlation
coefficient mean values (at lag=0) during supine and standing periods
were 0.88±0.05 and 0.85±0.06, respectively.
|
In contrast to baroreflex sensitivity values calculated by power spectral analysis, values calculated by CDM demonstrated dynamic variations in baroreflex sensitivity across time in all subjects. The magnitude of the time-dependent variations in baroreflex sensitivity was expressed quantitatively as the coefficient of variation. The coefficient of variation (100xSD/mean) was determined in each subject for each sampling period (supine and standing). A change in posture from supine to standing significantly decreased the mean coefficient of variation from 63.1±9.8% to 46.1±7.7% (P=.0001) in CDM analysis.
When baroreflex sensitivity was derived by sequential analysis
of spontaneous variations in pressure and RR interval, the magnitude of
the regression slope also varied over time (Fig 4
). A
change in posture also reduced the coefficient of variation using
sequential analysis from 51.9±12.1% to 40.7±7.7%
(P=.0001).
|
| Discussion |
|---|
|
|
|---|
Our objective in the present study was to determine the ability of CDM to provide a continuous assessment of baroreflex sensitivity in the time domain. The analysis of simulated data showed that CDM yielded the same values for baroreflex sensitivity as did power spectral analysis. The simulated data included a sudden change in baroreceptor sensitivity that was designed to simulate a change in posture from the supine to the standing position. The sudden change in baroreceptor sensitivity was reproduced by the CDM method with only minimal filter-induced ringing at the point of transition. This theoretical analysis suggests that CDM is capable of accurately assessing dynamic variations in baroreflex sensitivity. Furthermore, the time and frequency resolutions of CDM were found to be adequate to provide a dynamic assessment of baroreflex sensitivity over the short sampling period in the 33 subjects.
CDM of the actual data yielded mean values of baroreflex sensitivity
during supine and standing positions that were in agreement with
previously reported values in humans.7 8 9 10 11 12 13 In addition, the
mean values of baroreflex sensitivity derived by CDM in the present
study were similar to those obtained by power spectral analysis
(Fig 3
) and sequential analysis (Table
) on the same time
series. All three methods of analysis showed that a change in
posture from the supine to the standing position was accompanied by a
significant decrease in baroreflex sensitivity. Similar changes in
baroreflex sensitivity with a change in posture have been reported
previously in humans with the use of power spectral
analysis11 12 and sequential
analysis.12 A postural change from supine to
standing decreases cardiac output,21 increases sympathetic
vasomotor traffic,22 and decreases cardiac vagal efferent
tone.23 Ferrari et al24 showed that
sympathetic activity exerts an antagonistic influence on
the baroreceptor control of heart rate in conscious rats. Thus, it is
likely that postural-induced changes in autonomic tone are responsible
for the changes in baroreflex sensitivity observed in the present
study.
In contrast to power spectral analysis, CDM provided a
continuous assessment of baroreflex sensitivity in the time domain. The
results of CDM showed that baroreflex sensitivity decreased rapidly to
a steady-state value when the subjects assumed an upright posture.
Furthermore, in both the supine and upright postures, the gain of the
baroreflex was not static but demonstrated oscillations
over time. Time-dependent changes in baroreflex sensitivity could not
be discerned by power spectral analysis but could be detected
by sequential analysis. The dynamic characteristics of
baroreflex sensitivity were evident by variations in the slope of the
regression of systolic BP against RR interval (Fig 4
). Since
sequential analysis is a static method that can be applied only
during selected portions of the time series, it was not possible to
directly compare the results of sequential analysis and CDM in
real time. With the use of the coefficient of variation as an index of
the magnitude of time-dependent fluctuations in baroreflex sensitivity,
there was a 17±13% (CDM) or 12±8.8% (sequential analysis)
decrease in variability from the supine to standing position
(P=.0001). Although the mechanism for this decrease is not
entirely clear, it may be related to changes in autonomic tone upon
standing.
The results of the present study strongly indicate that baroreflex sensitivity is not static but fluctuates over time. However, the evidence obtained with simulated and actual data was indirect. Therefore, proper validation of CDM analysis should include its application to the analysis of BP and RR interval recordings obtained in experimental animals before and after sinoaortic denervation.
Baroreflex sensitivity was assessed from the analysis of only low-frequency oscillations in the present study. However, other investigators have suggested that the cross-spectral modulus of high-frequency oscillations (0.15 to 0.45 Hz) is also an index of baroreflex sensitivity.8 9 12 DeBoer et al25 proposed in their theoretical model that RR interval oscillations in the high-frequency band (respiratory sinus arrhythmia) were mainly determined by baroreflex-mediated vagal efferent activity. However, cardiac vagal efferent activity has also been shown to be modulated directly by an influence of medullary respiratory neurons on cardiovascular neurons or reflexly by changes in lung inflation.26 It is also possible that respiratory movement mechanically perturbs both BP and the RR interval,27 since the coherence between RR interval and BP oscillations in this frequency band does not disappear even after sinoaortic denervation.28 Furthermore, heart rate variations at the respiratory frequency do not appear to be mediated by changes in sympathetic efferent activity.29 Since RR interval oscillations in the low-frequency band are due to baroreceptor-mediated alterations in both sympathetic and vagal efferent activities,25 29 only the low-frequency components were analyzed in the present study.
In summary, the present study explored the possibility of a continuous assessment of baroreflex sensitivity using CDM of RR interval and systolic BP variabilities. The results show that baroreflex sensitivities derived by CDM are equivalent to those derived by power spectral analysis and sequential analysis. In addition, CDM has the capability of providing dynamic changes in baroreflex sensitivity as a function of time. This technique may be a useful tool in exploring dynamic changes in reflex autonomic control of the cardiovascular system in individuals with hypertension and other cardiac abnormalities.
| Footnotes |
|---|
| Appendix 1 |
|---|
|
|
|---|
![]() | (1) |
![]() | (2) |
t, and
t are the changing amplitudes and phases of the
frequency components of the RR interval and BP, respectively; and
MRt and MSt are residual time series including
all other components and noises such as direct current (DC or mean)
trends. The aim of CDM is to extract approximations of the amplitudes
and phases as a function of time (t).
The complex analogs of Equations 1
, and 2
are written as
![]() | (3) |
![]() |
![]() | (4) |
![]() |
![]() | (5) |
![]() |
![]() | (6) |
![]() |
![]() | (7) |
![]() | (8) |
![]() | (9) |
![]() | (10) |
Received April 17, 1996; first decision May 27, 1996; accepted September 17, 1996.
| References |
|---|
|
|
|---|
2.
Smyth HS, Sleight P, Pickering GW. Reflex
regulation of arterial pressure during sleep in
man. Circ Res.. 1969;24:109-121.
3.
Pickering TG, Gribbon B, Strange Petersen E,
Cunningham DJC, Sleight P. Effects of autonomic blockade on the
baroreflex in man at rest and during exercise. Circ
Res.. 1972;30:177-185.
4. Eckberg DL. Parasympathetic cardiovascular control in human disease: a critical review of methods and results. Am J Physiol.. 1980;239:H581-H593.
5.
Bertinieri G, Di Rienzo M, Cavallazzi A, Ferrari AU,
Pedotti A, Mancia G. Evaluation of baroreceptor reflex by blood
pressure monitoring in unanesthetized cats.
Am J Physiol.. 1988;254:H377-H383.
6.
Parati G, Di Rienzo M, Bertinieri G, Pomidossi
G, Casadei R, Groppelli A, Pedotti A, Zanchetti A, Mancia G.
Evaluation of the baroreceptor-heart rate reflex by 24-hour
intra-arterial blood pressure monitoring in humans.
Hypertension.. 1988;12:214-222.
7.
Robbe HWJ, Mulder LJM, Ruddel H, Langewitz WA, Veldman
JBP, Mulder G. Assessment of baroreceptor reflex sensitivity by
means of spectral analysis. Hypertension.. 1987;10:538-543.
8.
Pagani M, Somers V, Furlan R, Dell'Orto S, Conway J,
Baselli G, Cerutti S, Sleight P, Malliani A. Changes in
autonomic regulation in mild hypertension.
Hypertension.. 1988;12:600-610.
9. Hughson RL, Quintin L, Annat G, Yamamoto Y, Gharib C. Spontaneous baroreflex by sequence and power spectral methods in humans. Clin Physiol.. 1993;13:663-676.[Medline] [Order article via Infotrieve]
10.
Parati G, Mutti E, Frattola A, Castiglioni P, Di Rienzo
M, Mancia G. ß-Adrenergic blocking treatment and 24-hour
baroreflex sensitivity in essential hypertensive patients.
Hypertension.. 1994;23:992-996.
11.
Veerman DP, Imholz BPM, Wieling W, Karemaker JM, van
Montfrans GA. Effects of aging on blood pressure variability in
resting conditions. Hypertension.. 1994;24:120-130.
12. Munakata M, Imai Y, Takagi H, Nakao M, Yamamoto M, Abe K. Altered frequency-dependent characteristics of the cardiac baroreflex in essential hypertension. J Auton Nerv Syst.. 1994;49:33-45.[Medline] [Order article via Infotrieve]
13.
Parlow J, Viale J-P, Annat G, Hughson R, Quintin
L. Spontaneous cardiac baroreflex in humans: comparison with
drug-induced responses. Hypertension.. 1995;25:1058-1068.
14.
Di Rienzo M, Parati G, Castiglioni P, Omboni S, Ferrari
AU, Ramirez AJ, Pedotti A, Mancia G. Role of sinoaortic
afferents in modulating BP and pulse interval spectral characteristics
in unanesthetized cats. Am J Physiol.. 1991;261:H1811-H1818.
15. Bloomfield P. Complex demodulation. In: Fourier Analysis of Time Series: An Introduction. New York, NY: John Wiley & Sons; 1976:118-150.
16. Shin S-J, Tapp WN, Reisman SS, Natelson BH. Assessment of autonomic regulation of heart rate variability by the method of complex demodulation. IEEE Trans Biomed Eng.. 1989;36:274-283.[Medline] [Order article via Infotrieve]
17.
Hayano J, Taylor JA, Yamada A, Mukai S, Hori R,
Asakawa T, Yokoyama K, Watanabe Y, Takata K, Fujinami T.
Continuous assessment of hemodynamic control by complex
demodulation of cardiovascular variability.
Am J Physiol.. 1993;264:H1229-H1238.
18.
Hayano J, Taylor JA, Mukai S, Okada A, Watanabe Y,
Takata K, Fujinami T. Assessment of frequency shifts in R-R
interval variability and respiration with complex demodulation.
J Appl Physiol.. 1994;77:2879-2888.
19. Oppenheim AV, Schafer RW. Digital Signal Processing. Englewood Cliffs, NJ: Prentice Hall; 1975:155-556.
20.
Nearing BD, Verrier RL. Personal computer system
for tracking cardiac vulnerability by complex demodulation of the T
wave. J Appl Physiol.. 1993;74:2606-2612.
21. Smith JJ, Porth CJM. Posture and the circulation: the age effect. Exp Gerontol.. 1991;26:141-162.[Medline] [Order article via Infotrieve]
22. Rowell LB. Human Cardiovascular Control. Oxford, UK: Oxford University Press; 1993:1-162.
23.
Pomeranz B, Macaulay RJB, Caudill MA, Kutz I, Adam D,
Gordon D, Kilborn KM, Barger AC, Shannon DC, Cohen RJ, Benson H.
Assessment of autonomic function in humans by heart rate spectral
analysis. Am J Physiol.. 1985;248:H151-H153.
24.
Ferrari AU, Daffonchio A, Franzelli C, Mancia G.
Potentiation of the baroreceptorheart rate reflex by
sympathectomy in conscious rats.
Hypertension.. 1991;18:230-245.
25.
DeBoer RW, Karemaker JM, Strackee J.
Hemodynamic fluctuations and baroreflex sensitivity in
humans: a beat-to-beat model. Am J Physiol.. 1987;253:H680-H689.
26.
Hirsh JA, Bishop B. Respiratory sinus
arrhythmia in humans: how breathing pattern modulates heart
rate. Am J Physiol.. 1981;241:H620-H629.
27. Blinks JR. Positive chronotropic effect of increasing right atrial pressure in the isolated mammalian heart. Am J Physiol.. 1956;186:299-303.
28. Parati G, Omboni S, Frattola A, Di Rienzo M, Zanchetti A, Mancia G. Dynamic evaluation of the baroreflex in ambulant subjects. In: Di Rienzo M, Mancia G, Parati G, Pedotti A, Zanchetti A, eds. Blood Pressure and Heart Rate Variability. Amsterdam, Netherlands: IOS Press; 1992:123-137.
29.
Berger RD, Saul JP, Cohen RJ. Transfer function
analysis of autonomic regulation, I: canine atrial rate
response. Am J Physiol.. 1989;256:H142-H152.
This article has been cited by other articles:
![]() |
Y. C. Tzeng, P. D. Larsen, and D. C. Galletly Mechanism of cardioventilatory coupling: insights from cardiac pacing, vagotomy, and sinoaortic denervation in the anesthetized rat Am J Physiol Heart Circ Physiol, April 1, 2007; 292(4): H1967 - H1977. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Bauernschmitt, H. Malberg, N. Wessel, B. Kopp, E.U. Schirmbeck, and R. Lange Impairment of cardiovascular autonomic control in patients early after cardiac surgery Eur. J. Cardiothorac. Surg., March 1, 2004; 25(3): 320 - 326. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D. Lipman, J. K. Salisbury, and J. A. Taylor Spontaneous Indices Are Inconsistent With Arterial Baroreflex Gain Hypertension, October 1, 2003; 42(4): 481 - 487. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Sander, C. Kukla, J. Klingelhofer, K. Winbeck, and B. Conrad Relationship Between Circadian Blood Pressure Patterns and Progression of Early Carotid Atherosclerosis : A 3-Year Follow-Up Study Circulation, September 26, 2000; 102(13): 1536 - 1541. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Hypertension Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |