(Hypertension. 1999;33:726-731.)
© 1999 American Heart Association, Inc.
Scientific Contribution |
From the Department of Internal Medicine, Angiology Division (U.H., L.R., E.S.), and Institute of Diagnostic Radiology (D.A.L., T.P., T.F.H., P.H., G.G.Z.-P., J.F.D.), Zürich University Hospital, Zürich, Switzerland.
Correspondence to Jörg F. Debatin, MD, Institute of Diagnostic Radiology, Zürich University Hospital, Rämistr. 100, CH-8091 Zürich, Switzerland. E-mail debatin{at}dmr.usz.ch
| Abstract |
|---|
|
|
|---|
60% diameter reduction) main renal artery
stenosis, sensitivity and specificity were 90% and 86%,
respectively, for magnetic resonance angiography and 81% and 87% for
duplex sonography. Most false readings involved differential grading of
stenoses detected with all 3 techniques. When patients with
fibromuscular dysplasia were excluded from the analysis, the
sensitivity of magnetic resonance angiography increased to 97%, with a
negative predictive value of 98%. Magnetic resonance angiography
detected 96% and duplex 5% of accessory renal arteries seen at
catheter angiography. Contrast-enhanced magnetic resonance angiography
is a useful technique for diagnosing atherosclerotic renovascular
disease. It overcomes the major limitations of duplex renal scanning.
However, duplex has the advantage of providing
hemodynamic information and appears better suited for
the assessment of patients with suspected fibromuscular dysplasia.
Key Words: arteries hypertension, renovascular magnetic resonance imaging renal artery obstruction ultrasound
| Introduction |
|---|
|
|
|---|
Conventional catheter angiography remains the gold standard for the diagnosis of renovascular disease. However, the relatively high complication rate and cost associated with catheter arteriography preclude its use as a primary diagnostic measure. A less invasive screening test is therefore desirable.14 15 Color-flow duplex sonography of the renal arteries has been shown to be a useful noninvasive method for the detection of renal artery stenosis16 17 18 and is a well-established technique in many institutions, including our own. On the basis of differential flow velocity in the aorta and renal artery, the hemodynamic significance of a renal artery lesion is functionally assessed. Duplex renal scanning is limited, however, by an inherent operator-dependence, a technical failure rate ranging from 10% to 20%, and the inability to visualize accessory renal arteries.16 17 18
Because of its flow sensitivity, magnetic resonance imaging (MRI) has also been used for noninvasive depiction of the renal arteries.19 20 21 However, lengthy imaging times, inconsistent visualization of distal renal artery segments, and difficulty in grading stenoses have remained barriers to the clinical acceptance of conventional magnetic resonance angiography (MRA). Recent advances in magnetic resonance hardware have allowed the development of a new approach to MRA whereby rapid breath-hold data acquisition is combined with dynamic contrast medium infusion.22 23 24 The aim of the current study was to prospectively assess the diagnostic value of breath-hold three-dimensional (3D) contrast-enhanced MRA of the renal arteries compared with color-flow renal duplex scanning in patients with clinical suspicion of renovascular disease using conventional catheter angiography as the standard of reference.
| Methods |
|---|
|
|
|---|
|
All patients underwent MRA and duplex sonography. Sixty patients underwent catheter arteriography of the renal arteries. These included all patients with pathological findings at either duplex or MRA, as well as patients with concomitant peripheral vascular disease who underwent conventional angiographic evaluation of the abdominal aorta and lower extremities. All studies were performed in accordance with the guidelines set forth by the institutional review board. Written informed consent was obtained from all patients.
MR Imaging
MR examinations were conducted on a 1.5-T scanner (Signa Horizon
Echospeed, GE Medical Systems) using a phased-array surface coil (Torso
coil) for signal reception. The pulse sequence used for MRA was a
coronal spoiled gradient-echo acquisition with a repetition time of 4.7
milliseconds, an echo time of 2.1 milliseconds, a flip angle of 50°,
and a matrix of 256x160. Partition thickness and field of view were
adjusted to patient size and ranged from 1.6 to 2.0 mm and 30 to
38 cm, respectively. The data were acquired in a breath-hold interval
lasting 27 seconds timed to coincide with the arterial
phase of a dynamic bolus infusion of 0.3 mmol/kg body wt
gadolinium DTPA at a rate of 3.0 mL/s. The timing protocol was based on
determination of the contrast travel time from the antecubital venous
injection site to the aorta using a 2-mL test bolus.25
Duplex Sonography
All renal duplex studies were performed by the same vascular
technologist using a duplex scanner with a 3.5-MHz transducer (Acuson
128 XP-10, Acuson Corp). Doppler velocity waveforms were obtained
along the entire course of the renal arteries as well as the abdominal
aorta at the level of the renal arteries. Duplex studies were
considered diagnostic if the entire renal artery could be
interrogated by velocity waveform analysis. A focal increase in
the angle-adjusted peak systolic velocity in a renal artery
indicated stenosis. If this was followed by turbulence and the
ratio of maximum peak systolic velocity in the renal artery to
the peak systolic velocity in the adjacent aorta was >3.5, the
stenosis was considered hemodynamically
significant (
60% diameter reduction). Renal artery occlusion was
diagnosed by the lack of flow signal in a visualized renal artery
associated with low amplitude velocity signal from the renal parenchyma
and a small kidney (<9 cm).
Duplex scans with diagnosis of hemodynamically significant renal artery stenosis where the contralateral renal artery was not adequately visualized (n=3) were considered diagnostic examinations because the patient would proceed to conventional angiography on the basis of that finding alone.
Conventional Angiography
Catheter arteriography was performed in digital subtraction or
cut film technique (Siemens Digitron 3). Initially, an anteroposterior
abdominal aortogram was obtained with a transfemorally inserted 5F
pigtail catheter positioned at the level of the renal arteries.
Multiple injections at different projections, magnification views,
and selective runs were performed as required.
Image Interpretation and Statistical Analysis
All examinations were interpreted blinded to the results of the
other diagnostic tests. With all 3 modalities, main and
accessory renal arteries were graded on a 4-point scale as (1) normal,
(2) not significantly stenosed (
60%), (3) significantly stenosed
(
60%), or (4) occluded. Renal arteries with evidence of
fibromuscular dysplasia were graded as significantly stenosed (grade 3)
because it was assumed that such lesions are
hemodynamically significant in patients with clinical
findings suggestive of renovascular disease.
Duplex scans were interpreted by the same vascular technologist performing the examination. Magnetic resonance images and conventional angiograms were interpreted independently by 2 magnetic resonance radiologists and 2 vascular radiologists, respectively. With both techniques, discordant cases were reviewed by the 2 interpreters so a consensus could be reached.
With conventional angiography and MRA, the examining radiologist simply recorded the observable decrease in luminal diameter. Magnetic resonance images were interpreted at a console with the postprocessing software provided by GE Medical Systems. Postprocessing techniques used for MRA included maximum intensity projections and 3D multiplanar reformations. The former is an algorithm that allows display of a 3D volume of image data as a two-dimensional projection and renders images similar to those of conventional angiography. 3D multiplanar reformations, the more important technique, allows the user to view the 3D imaging volume in cross section in any desired plane.
Sensitivity, specificity, positive and negative predictive values, and
two-sided 95% confidence intervals of MRA and duplex sonography
regarding the detection of significant main renal artery
stenosis were calculated. For these calculations, renal
arteries graded as normal or not significantly stenosed were considered
negative findings, and those graded as significantly stenosed or
occluded were considered positive findings. The statistical
significance of the differences in sensitivities between MRA and duplex
were assessed by means of the McNemar test.26
Interobserver variability in the interpretation of conventional
angiography and MRA were computed using the
statistic including
95% confidence intervals.
| Results |
|---|
|
|
|---|
Thirty-four patients had positive findings for significant renal artery stenosis confirmed by conventional angiography, (Figures 1 and 2) including 5 renal arteries in 4 patients with fibromuscular dysplasia (bilateral disease in 1 patient) and 40 renal arteries in 30 patients with atherosclerosis (bilateral disease in 10 patients). A total of 60 patients containing 119 kidneys underwent catheter arteriography. The varying number of nondiagnostic studies with duplex and MRA resulted in a total of 85 main renal arteries that had a reading with all 3 tests. A correlation between MRA and conventional angiography was available in 109 main renal arteries, and a correlation between duplex and conventional angiography was available in 90 main renal arteries. With MRA, 10 false-positive and 4 false-negative readings were recorded. Three of the 4 false-negative findings involved renal arteries with fibromuscular dysplasia, and 1 was a borderline atherosclerotic lesion that was graded as significant at conventional angiography and not significant at MRA. Both of these readings were determined by consensus opinion following divergent readings by the individual reviewers. The 10 false-positive MRA readings were overestimations of stenoses graded at conventional angiography to be not significantly stenosed. Three of the 10 lesions were also graded as significant stenoses at duplex.
|
|
With duplex ultrasound, 7 false-positive and 7 false-negative readings for significant renal artery stenosis were recorded. Five of the 7 false-positive readings involved atherosclerotic lesions that were interpreted as significant with duplex and not significant at conventional angiography. Four of these were graded as significant stenoses following consensus reading at conventional angiography, and 3 of these were graded as significant stenoses at MRA. Of the 7 false-negative readings with duplex, 1 was a renal artery with discrete distal fibromuscular dysplasia that was also missed by MRA, and 3 were stenoses graded as not significant at duplex and significant at conventional angiography and MRA.
Sensitivity, specificity, positive and negative predictive values for
duplex sonography, and MRA are shown in Table 2 along with probability value of the
McNemar statistic for comparison of differences in sensitivities
between MRA and duplex. The interobserver variability in the grading of
stenotic renal artery lesions (grades 1 to 4) with conventional
angiography and MRA were identical, with a
value of 0.77 and 95%
confidence intervals ranging from 0.67 to 0.86. For the detection of
hemodynamically significant lesions, interobserver
variability was 0.87 (0.78 to 0.95) for MRA and 0.88 (0.79 to 0.97) for
conventional angiography.
|
Multiple renal arteries were present in 16 patients (27%). Twenty-four of 25 (96%) accessory renal arteries detected with conventional angiography were also seen with MRA, including 2 significant stenoses and 20 normal arteries that were correctly graded. One accessory renal artery with a nonsignificant stenosis and another with a significant stenosis at conventional angiography were falsely graded by MRA as significant and not significant, respectively. Both were extremely small arteries. With duplex scanning, 1 of 20 (5%) accessory renal arteries was detected and correctly graded as normal.
| Discussion |
|---|
|
|
|---|
Breath-hold contrast-enhanced MRA appears to overcome the limitations of duplex renal scanning. The examination is fast, is operator-independent, is reliable in the detection of accessory renal arteries, and has a low technical failure rate. In clinical practice, the problem posed by claustrophobia can, at least in part, be mitigated by the use of sedatives. Since arterial contrast is derived from the presence of a contrast agent rather than being flow-dependent, contrast-enhanced MRA does not suffer from the saturation- and turbulence-induced signal loss inherent to conventional noncontrast MRA.19 20 21 Indeed, the results of this study show good correlation between contrast-enhanced MRA and conventional arteriography for grading stenotic renal artery lesions. When patients with fibromuscular dysplasia were excluded from the analysis, the sensitivity was 97% and the negative predictive value 98%. This low rate of false-negative findings demonstrates the potential usefulness of MRA as a screening test for renal artery stenosis in a patient population with clinical suspicion of renovascular disease. The McNemar test, which assesses the statistical significance of the observed differences in the sensitivities between MRA and duplex, yielded a P value of 0.125 when patients with fibromuscular dysplasia were excluded. This represents a statistical trend suggesting that the two techniques are not equally sensitive, thus favoring MRA over duplex in this patient subset.
Contrast-enhanced MRA is similar to conventional angiography in that
both techniques provide a luminogram of arterial
morphology. Grading of stenoses with both techniques is based
on measuring the amount of luminal narrowing. The similarity between
the 2 techniques is mirrored in the comparable
values for
interobserver variability. However, in its present form,
contrast-enhanced MRA has a spatial resolution inferior to
that of catheter angiography. This is reflected in the poor
performance of MRA in the present study regarding diagnosis
of fibromuscular dysplasia. MRA is a rapidly evolving technique,
however, and improvements in both spatial and temporal resolutions can
be expected in the future. Moreover, since 3D multiplanar reformations
allow visualization of the MR data in operator-defined cross-sectional
planes, MRA does not share the projection-related limitations of
conventional arteriography.27 This facilitates depiction
of eccentric lesions and permits reliable localization of renal artery
origins in patients with vascular tortuosity. Although the acquisition
of MRA examinations is operator-independent, considerable expertise is
required for effective postprocessing and interpretation of the images.
The morphological depiction and display offered by MRI
represents an advantage of MRA over duplex ultrasound, not only
for therapeutic planning purposes but also for the assessment of
additional soft tissue lesions. One patient from this study was
diagnosed to have renal cell carcinoma at MRA, which had been missed at
duplex scanning.
Several methodological limitations of this study need to be considered. Patients with nondiagnostic duplex examinations as well as claustrophobic patients unable to undergo MRA were excluded from the modality-related statistical analysis, thereby possibly introducing a selection bias. Furthermore, only 60 of the 89 enrolled patients underwent conventional angiography. Although use of conventional angiography as the standard of reference for the entire study population would have been preferable, ethical concerns regarding potential risks of catheter angiography led to the exclusion of those patients with concordant negative MRA and duplex sonography studies. This may have contributed to a possible underestimation of false-negative findings. Finally, accessory renal arteries were not included in the sensitivity and specificity analysis because most were missed with duplex sonography. With MRA, on the other hand, 24 of 25 accessory vessels were visualized.
Similar to conventional angiography, the major drawback of contrast-enhanced MRA is the reliance on purely morphological criteria for the diagnosis of renovascular disease. The advantage of duplex sonography is that it assesses the hemodynamic effects of renal artery lesions. Although with inferior temporal resolution compared with duplex, accurate renal flow velocity measurements are possible with motion-sensitive MR pulse sequences.28 A recent study reported high sensitivity and specificity for detection of renal artery stenosis using MR flow velocity profiles.29 Lack of functional data may also be implicated as the cause for the suboptimal performance of conventional angiography in a recent study comparing selective catheter angiography with intraarterial pressure measurements for the assessment of hemodynamically significant renal artery stenosis.30 So despite its exquisite spatial resolution, catheter angiography is by no means an absolute measure of hemodynamic significance. Hence, the results of the present study must be interpreted with caution because most of the false readings with MRA as well as duplex involved differential grading of stenoses detected with all 3 techniques. Indeed, the hemodynamic significance of a borderline stenosis seen at MRA and conventional angiography may best be evaluated by measuring the flow alteration with duplex sonography.
For a diagnostic test to be useful in routine clinical practice, particularly as a screening method, it needs to be available, cost-effective, and safe as well as diagnostically accurate. Contrast-enhanced MRA requires the intravenous administration of a contrast agent. In contrast to iodinated contrast agents, gadolinium is not nephrotoxic and can thus be used in patients with renal insufficiency without the risk of further impairment of renal function.31 32 Both the availability and cost of MRA are less suitable than those of duplex sonography; however, considering the substantial rate of nondiagnostic examinations as well as the lengthy examination times of duplex compared with MRA, the ultimate cost-effectiveness of the 2 techniques regarding the detection of renovascular disease will need to be addressed in future studies.
In conclusion, our results demonstrate that breath-hold contrast-enhanced MRA is a useful technique for the detection of atherosclerotic renal artery stenosis. Although the spatial resolution appears inadequate for consistent detection of fibromuscular dysplasia, improvements can be expected in the future. Contrast-enhanced MRA is fast, operator-independent, and reliable in the detection of accessory renal arteries and thus overcomes most of the limitations inherent in duplex sonography. Duplex, on the other hand, has the advantage of providing functional flow information and is therefore useful for assessing hemodynamic significance. At present, we can recommend MRA for the morphological assessment of suspected atherosclerotic renal artery stenosis, particularly in obese patients who are not amenable to ultrasound examination. However, duplex appears to be better suited for the assessment of patients with suspected fibromuscular dysplasia.
| Acknowledgments |
|---|
Received June 2, 1998; first decision July 20, 1998; accepted November 2, 1998.
| References |
|---|
|
|
|---|
2. Missouris CG, Buckenham T, Cappuccio FP, MacGregor GA. Renal artery stenosis: a common and important problem in patients with peripheral vascular disease. Am J Med. 1994;96:1014.[Medline] [Order article via Infotrieve]
3. Harding MB, Smith LR, Himmelstein SI, Harrison K, Phillips HK, Schwab SJ, Hermiller JB, Davidson CJ, Bashore TM. Renal artery stenosis: prevalence and associated risk factors in patients undergoing cardiac catheterization. J Am Soc Nephrol. 1992;88:46N50N.
4.
Working Group on Renovascular Hypertension. Detection,
evaluation and treatment of renovascular hypertension. Arch
Intern Med. 1987;147:820829.
5. Textor SC. Renovascular hypertension. Endocrinol Metab Clin North Am. 1994;23:227234.
6. Jacobson HR. Ischemic renal disease: an overlooked clinical entity? Kidney Int.. 1988;34:729743.[Medline] [Order article via Infotrieve]
7. O'Neil EA, Hansen KJ, Canzanello VJ, Pennell TC, Dean RH. Prevalence of ischemic nephropathy in patients with renal insufficiency. Am Surg. 1992;58:485490.[Medline] [Order article via Infotrieve]
8. Schreiber MJ, Pohl MA, Novick AC. The natural history of atherosclerotic and fibrous renal artery disease. Urol Clin North Am. 1984;91:881887.
9. Zierler RE, Bergelin RO, Isaacson JA, Strandness DE. Natural history of atherosclerotic renal artery stenosis: a prospective study with duplex ultrasonography. J Vasc Surg. 1994;19:250258.[Medline] [Order article via Infotrieve]
10.
Dean RH, Kieffer RW, Smith BM. Renovascular
hypertension: anatomic and renal function changes during drug therapy.
Arch Surg. 1981;116:14081415.
11. Sos TA, Pickering TG, Sniderman K, Saddekni S, Case DB, Silane MF, Vaughan ED Jr, Laragh JH. Percutaneous transluminal renal angioplasty in renovascular hypertension due to atheroma or fibromuscular dysplasia. N Engl J Med. 1983;309:274279.[Abstract]
12. Ying CY, Tifft TP, Gavaras H, Chobanian AV. Renal revascularization in the azotemic hypertensive patient resistant to therapy. N Engl J Med. 1984;311:10701075.[Abstract]
13. Weibull H, Bergquist D, Bergentz SE, Jonsson K, Hulthen L, Manhem P. Percutaneous transluminal renal angioplasty versus surgical reconstruction of atherosclerotic renal artery stenosis: a prospective randomized study. J Vasc Surg. 1993;18:841850.[Medline] [Order article via Infotrieve]
14. Mann SJ, Pickering TG. Detection of renovascular hypertension. Ann Intern Med. 1992;117:845853.
15. Canzanello VJ, Textor SC. Noninvasive diagnosis of renovascular disease. Mayo Clin Proc. 1994;69:11721181.[Medline] [Order article via Infotrieve]
16. Taylor DC, Kettler MD, Moneta GL. Duplex ultrasound scanning in the diagnosis of renal artery stenosis: a prospective evaluation. J Vasc Surg. 1986;7:363369.
17. Hoffmann U, Edwards JM, Carter S. Role of duplex scanning for the detection of atherosclerotic renal artery disease. Kidney Int. 1991;39:12321239.[Medline] [Order article via Infotrieve]
18.
Olin JW, Piedmonte MR, Young JR, DeAnna S, Grubb M,
Childs MB. The utility of duplex ultrasound of the renal arteries for
diagnosing significant renal artery stenosis. Ann Intern
Med. 1995;122:833838.
19.
Kim D, Edelman RR, Kent KC, Porter DH, Skillman JJ.
Abdominal aorta and renal artery stenosis: evaluation with MR
angiography. Radiology. 1990;174:727731.
20.
Debatin JF, Spritzer CE, Grist TM, Beam C, Svetkey CP,
Newman GE, Sostman HD. Imaging of the renal arteries: value of MR
angiography. Am J Roentgenol. 1991;157:981990.
21.
De Cobelli F, Mellone R, Salvioni M, Vanzulli A, Sironi
S, Manunta P, Lanzani C, Bianchi G, Del Maschio A. Renal artery
stenosis: value of screening with three-dimensional
phase-contrast MR angiography with a phased-array multicoil.
Radiology. 1996;201:697703.
22.
Prince MR. Gadolinium-enhanced MR aortography.
Radiology. 1994;191:155164.
23.
Prince MR, Narasimhan DL, Stanley JC, Chenevert TC,
Williams DM, Marx MV, Cho KJ. Breath-hold gadolinium-enhanced MR
angiography of the abdominal aorta and its major branches.
Radiology. 1995;197:785792.
24.
Leung DA, McKinnon GC, Davis CP, Pfammatter T, Krestin
GP, Debatin JF. Breath-hold contrast-enhanced three-dimensional MR
angiography. Radiology. 1996;201:569571.
25. Hany TF, McKinnon GC, Leung DA, Pfammatter T, Debatin JF. Optimization of contrast timing for breath-hold three-dimensional MR angiography. J Magn Reson Imaging. 1997;7:551556.[Medline] [Order article via Infotrieve]
26.
Dwyer AJ. Matchmaking and McNemar in the comparison of
diagnostic modalities. Radiology. 1991;178:328330.
27.
Foster JH, Klatte EC, Burko H. Arteriographic pitfalls
in the diagnosis of renovascular hypertension. Arch Surg. 1969;99:792801.
28.
Debatin JF, Ting RH, Wegmüller H, Sommer FG,
Fredrickson JD, Brosnan TJ, Bowman BS, Myers BD, Herfkens RJ, Pelc NJ.
Renal artery blood flow: quantitation with phase-contrast MR imaging
with and without breath holding. Radiology. 1994;190:371378.
29.
Schoenberg SO, Knopp MV, Bock M, Kallinowski F, Just A,
Essig M, Hawighorst H, Schad L, van Kaick G. Renal artery
stenosis: grading of hemodynamic changes with
cine phase-contrast MR blood flow measurements. Radiology. 1997;203:4553.
30.
Wasser MN, Westenberg J, van der Hulst VP, van
Baelen J, van Bockel JH, van Erkel AR, Pattynama PM.
Hemodynamic significance of renal artery
stenosis: digital subtraction angiography versus
systolically gated three-dimensional phase-contrast MR
angiography. Radiology. 1997;202:333338.
31. Niendorf HP, Haustein J, Cornelius I, Alhassan A, Clauss W. Safety of gadolinium-DTPA: extended clinical experience. Magn Reson Med. 1991;22:222228.[Medline] [Order article via Infotrieve]
32. Prince MR, Arnoldus C, Frisoli JK. Nephrotoxicity of high-dose gadolinium compared with iodinated contrast. J Magn Reson Imaging. 1996;6:162166.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
B. Drieghe, J. Madaric, G. Sarno, G. Manoharan, J. Bartunek, G. R. Heyndrickx, N. H.J. Pijls, and B. De Bruyne Assessment of renal artery stenosis: side-by-side comparison of angiography and duplex ultrasound with pressure gradient measurements Eur. Heart J., February 2, 2008; 29(4): 517 - 524. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. J. Williams, P. Macaskill, S. F. Chan, T. E. Karplus, W. Yung, E. M. Hodson, and J. C. Craig Comparative Accuracy of Renal Duplex Sonographic Parameters in the Diagnosis of Renal Artery Stenosis: Paired and Unpaired Analysis Am. J. Roentgenol., March 1, 2007; 188(3): 798 - 811. [Abstract] [Full Text] [PDF] |
||||
![]() |
S J Gandy, K Armoogum, R S Nicholas, T B McLeay, and J G Houston A clinical MRI investigation of the relationship between kidney volume measurements and renal function in patients with renovascular disease Br. J. Radiol., January 1, 2007; 80(949): 12 - 20. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Nael, R. Saleh, M. Lee, T. McNamara, S. R. Godinez, G. Laub, J. P. Finn, and S. G. Ruehm High-spatial-resolution contrast-enhanced MR angiography of abdominal arteries with parallel acquisition at 3.0 T: initial experience in 32 patients. Am. J. Roentgenol., July 1, 2006; 187(1): W77 - W85. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. E. Zierler Atherosclerotic Renovascular Disease: Natural History and Diagnosis Perspectives in Vascular Surgery and Endovascular Therapy, December 1, 2004; 16(4): 299 - 310. [Abstract] [PDF] |
||||
![]() |
C. K. Guill, D. C. Benavides, C. Rees, A. Z. Fenves, and E. C. Burton Fatal Mesenteric Fibromuscular Dysplasia: A Case Report and Review of the Literature Arch Intern Med, May 24, 2004; 164(10): 1148 - 1153. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Gowda, A. L. Loeb, L. J. Crouse, and P. H. Kramer Complementary roles of color-flow duplex imaging and intravascular ultrasound in the diagnosis of renal artery fibromuscular dysplasia: Should renal arteriography serve as the "gold standard"? J. Am. Coll. Cardiol., April 16, 2003; 41(8): 1305 - 1311. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. G. Turi and M. R. Jaff Renal artery stenosis: searching for the algorithms for diagnosis and treatment J. Am. Coll. Cardiol., April 16, 2003; 41(8): 1312 - 1315. [Full Text] [PDF] |
||||
![]() |
J. K. Willmann, S. Wildermuth, T. Pfammatter, J. E. Roos, B. Seifert, P. R. Hilfiker, B. Marincek, and D. Weishaupt Aortoiliac and Renal Arteries: Prospective Intraindividual Comparison of Contrast-enhanced Three-dimensional MR Angiography and Multi-Detector Row CT Angiography Radiology, March 1, 2003; 226(3): 798 - 811. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. M. Bossuyt, J. B. Reitsma, D. E. Bruns, C. A. Gatsonis, P. P. Glasziou, L. M. Irwig, D. Moher, D. Rennie, H. C.W. de Vet, and J. G. Lijmer The STARD Statement for Reporting Studies of Diagnostic Accuracy: Explanation and Elaboration Ann Intern Med, January 7, 2003; 138(1): W1 - W12. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. M. Bossuyt, J. B. Reitsma, D. E. Bruns, C. A. Gatsonis, P. P. Glasziou, L. M. Irwig, D. Moher, D. Rennie, H. C.W. de Vet, and J. G. Lijmer The STARD Statement for Reporting Studies of Diagnostic Accuracy: Explanation and Elaboration Clin. Chem., January 1, 2003; 49(1): 7 - 18. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Spinosa, A. H. Matsumoto, J. F. Angle, K. D. Hagspiel, D. Cage, E. A. Bissonette, K. G. Koenig, C. R. Ayers, and K. McConnell Safety of CO2- and Gadodiamide-Enhanced Angiography for the Evaluation and Percutaneous Treatment of Renal Artery Stenosis in Patients with Chronic Renal Insufficiency Am. J. Roentgenol., May 1, 2001; 176(5): 1305 - 1311. [Abstract] [Full Text] [PDF] |
||||
![]() |
R G Woolfson Renal failure in atherosclerotic renovascular disease: pathogenesis, diagnosis, and intervention Postgrad. Med. J., February 1, 2001; 77(904): 68 - 74. [Full Text] |
||||
![]() |
S. D. Sarkar, D. N. Siegel, G. Soulez, and V. L. Oliva Invited Commentary Authors' Response RadioGraphics, September 1, 2000; 20(5): 1368 - 1372. [Full Text] [PDF] |
||||
![]() |
D. Weishaupt, S. G. Ruhm, C. A. Binkert, M. Schmidt, M. A. Patak, F. Steybe, S. McGill, and J. F. Debatin Equilibrium-Phase MR Angiography of the Aortoiliac and Renal Arteries Using a Blood Pool Contrast Agent Am. J. Roentgenol., July 1, 2000; 175(1): 189 - 195. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. G. Ruehm, T. F. Hany, T. Pfammatter, E. Schneider, M. Ladd, and J. F. Debatin Pelvic and Lower Extremity Arterial Imaging: Diagnostic Performance of Three-Dimensional Contrast-Enhanced MR Angiography Am. J. Roentgenol., April 1, 2000; 174(4): 1127 - 1135. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Hypertension Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1999 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |