(Hypertension. 2003;41:1016.)
© 2003 American Heart Association, Inc.
Scientific Contributions |
From the Department of Clinical Pharmacology, St Thomas Hospital, Centre for Cardiovascular Biology and Medicine, Kings College, London UK.
Correspondence to Dr P.J. Chowienczyk, Department of Clinical Pharmacology, St Thomas Hospital, Lambeth Palace Road, London SE1 7EH, UK. E-mail phil.chowienczyk{at}kcl.ac.uk
| Abstract |
|---|
|
|
|---|
Key Words: aorta hypertension, arterial pulse
| Introduction |
|---|
|
|
|---|
|
Applanation tonometry provides a noninvasive means for recording high-fidelity arterial pressure,6 and this methodology can be used to estimate aortic AIx. The most widely used approach is to perform radial artery tonometry and then apply a transfer function to calculate the aortic pressure waveform from the radial waveform.79 A similar approach can be used to derive aortic AIx from the carotid pulse. Owing to the proximity of the carotid artery to the aorta, the carotid waveform is modified little by the carotid-to-aorta transfer function and, even without the use of a transfer function, can be used to estimate aortic AIx.10,11 Radial tonometry is favored, however, because it is simple to perform and well tolerated. Carotid tonometry, on the other hand, requires a high degree of technical expertise11 to achieve optimal results and does cause minor discomfort. Although use of a radial-to-aortic transfer function for the measurement of central systolic blood pressure (which differs from brachial systolic blood pressure as a result of peripheral amplification) has been established,9,12 the accuracy of this approach for the determination of aortic AIx has been disputed. Aortic AIx depends on higher-frequency information than does systolic blood pressure. The transfer function shows greater between-subject variability at high frequencies8,9 and is therefore, less likely to provide an accurate estimate of aortic AIx than of systolic blood pressure. Previous studies have shown greater variation in the determination of aortic AIx than of central systolic blood pressure.9,13
These reservations regarding the transfer function approach prompted us to reevaluate its accuracy and to examine whether information similar to that provided by the aortic AIx can be obtained directly from the radial pulse without the use of a transfer function. That the timing and amplitude of pressure wave reflection influence the radial pulse has long been acknowledged,2 and a radial AIx (Figure 1) has been described,2 which might contain information similar to the aortic AIx. We examined the agreement between aortic AIx obtained by applying a transfer function to the radial pulse and aortic AIx obtained from the carotid pulse, with and without the use of a transfer function. We then examined the relation of the aortic AIx (derived from the transformed radial pulse) and the radial AIx derived without the use of a transfer function. We studied healthy subjects and older subjects with hypertension and coronary artery disease to obtain representative samples from groups expected to have low and high values of AIx. In a subgroup of healthy subjects, we administered vasoactive drugs to determine whether the changes in aortic AIx tracked those in the radial AIx.
| Methods |
|---|
|
|
|---|
-adrenoreceptor antagonists (37%), ß-adrenoreceptor antagonists (27%), angiotensin-converting enzyme inhibitors (27%), and angiotensin II receptor antagonists (17%). In the patients with CAD, treatment included ß-adrenoreceptor antagonists (95%), nitrates (58%), calcium-channel blockers (25%), diuretics (25%), angiotensin-converting enzyme inhibitors (17%), and
-adrenoreceptor antagonists (8%). Measurements were performed while subjects were supine in a quiet, temperature-controlled environment after at least 15 minutes of rest. Radial and carotid arterial pressure waveforms were obtained by applanation tonometry (Millar Instruments) with Sphygmocor software version 6.3 (Atcor). Three successive recordings were obtained from each artery, and measurements were repeated when the waveform(s) did not pass the automatic quality controls specified by the Sphygmocor software. Brachial artery blood pressure was measured in triplicate with an oscillometric device (Omron 705CP, Omron) and was used to calibrate the radial pressure pulse. The carotid pressure pulse was calibrated by using the diastolic and mean arterial pressures obtained from the brachial artery measurements. This method for calibrating the carotid pulse to provide an accurate aortic systolic blood pressure has been previously validated.14 The aortic pulse was synthesized from the radial pulse and from the carotid pulse by using the transfer functions supplied by Sphygmocor. Effects of vasoactive drugs were assessed in a subset of 12 of the healthy volunteers. A similar protocol was used to obtain carotid and radial pressure pulse recordings at baseline and during the last 10 minutes of a 15-minute intravenous infusion of norepinephrine (75 ng · kg-1 · min-1, n=8) or nitroglycerine (100 µg · min-1, n=4).
Statistics
Agreement between estimates of the same measurement was assessed by Bland-Altman plots. The mean difference between measurements was used to quantify systematic error and the SD of the differences, the variation in agreement.15 Pearson correlation coefficients are also presented, with the recognition that these are of limited value in the comparison of closely related measurements.15 Different measurements (aortic and radial AIx) were compared by regression analysis. A Bland-Altman plot was used to compare aortic AIx with aortic AIx predicted from radial AIx (by using the coefficients of the regression equation of radial vs aortic AIx). P<0.05 was considered significant. All tests were 2-tailed.
| Results |
|---|
|
|
|---|
|
Aortic AIx
Mean values of aortic AIx estimated from the transformed radial waveform were 8.3±8.7, 27.9±10.6, and 29.5±11.8% in the control, hypertensive, and CAD patients, respectively. Agreement between aortic AIx estimated from the transformed radial and transformed carotid waveforms was relatively poor (Figure 3), with correlation coefficients of 0.47, 0.83, and 0.84 and mean differences of -3.8±12.4, -8.8±6.0, and -5.8±7.1 mm Hg for the control, hypertensive, and CAD groups, respectively. The overall correlation coefficient for all groups was 0.84 and mean difference, -6.2±9.2%. The SD of the difference (2SD=18.4%) was substantial in comparison with the difference of
20% seen between mean values of AIx in the control and hypertensive/CAD patients. Similar results were obtained when aortic AIx obtained from the transformed radial pulse was compared with aortic AIx estimated directly from the carotid artery without using a transfer function (overall correlation coefficient, 0.86; 2SD of the difference, 22.6%).
|
Comparison of Aortic With Radial AIx
There was a close correlation (R>0.92 within each group and R=0.96 for all groups; Figure 4) between aortic AIx obtained by applying a transfer function to the radial waveform and the radial AIx obtained without use of a transfer function. For comparison with previous data, a Bland-Altman plot was constructed for the difference between aortic AIx and AIx estimated from the regression line relating aortic to radial AIx. 2SDs of the difference were 5.4%, less than that seen for the difference between aortic AIx derived from radial and carotid waveforms by using transfer functions. Changes in aortic AIx produced by nitroglycerine and norepinephrine were correlated with those in radial AIx (R=0.93, P<0.0001; Figure 5).
|
|
| Discussion |
|---|
|
|
|---|
14%, for example, is seen in association with 10 years of ageing.16 This discrepancy between values of AIx derived by applying transfer functions to the carotid and radial arteries suggests that one or the other or both of the transfer functions are of limited accuracy in predicting aortic Aix, especially in healthy subjects. Because the carotid artery is close to the aorta, carotid and aortic waveforms are similar, and the carotid-to-aortic transfer function is close to unity. Reconstruction of the aortic waveform from the carotid waveform is thus less sensitive to the exact form of the transfer function. Indeed, the aortic AIx can be estimated from the carotid without use of a transfer function.11 We found a similar discrepancy between aortic AIx derived by application of a transfer function to the radial waveform and aortic AIx estimated from the carotid, irrespective of whether a transfer function was applied to the carotid waveform. The majority of the error is thus most likely to reside in the radial-to-aortic transfer function. The discordance in accuracy with which central systolic blood pressure and aortic AIx can be derived from the radial-to-aortic transfer function is consistent with the characteristics of the transfer function. The transfer function is a "generalized transfer function" calculated from transfer functions from individual subjects. It is thus possible to calculate the 95% confidence intervals that express the interindividual spread of the transfer function. Systolic blood pressure is determined by the coefficients for the lower frequencies, and these show much less spread than do those for the higher frequencies that influence AIx. The effect of this is illustrated in Figure 6: when a radial-to-aortic transform is applied to a typical radial waveform by using the upper and lower 95% confidence limits for the coefficients, there is little variation in systolic pressure but marked variation in the aortic AIx.
|
Whereas use of a transfer function is required to determine central systolic blood pressure from the radial artery, it might not be necessary to assess central pressure wave reflection. That the form of the radial pulse is determined largely by the systemic circulation has been acknowledged since Frederick Mahomed17 developed the sphygmograph in 1872, and the influence of pressure wave reflection within the systemic circulation on the radial waveform has been emphasized by ORourke and colleagues (Kelly et al16). The whole concept of a transfer function is that the influence of the upper limb on radial pulse (however complex with regard to differential effects on the various frequency components) differs little between subjects. A transfer function can neither add nor subtract information contained within the radial waveform, and therefore, it should be possible to infer central pressure wave reflection directly from the radial waveform as well as from the transformed waveform. In the present study, we observed a close correlation (R>0.92 in all groups) between the aortic AIx derived by application of a transfer function to the radial waveform and the radial AIx calculated directly from the radial waveform without use of a transfer function. Indeed, the correlation between these indices was greater than that between values of aortic AIx derived from radial and carotid waveforms by using the transfer function approach. Furthermore, when changes in AIx were induced by vasoactive drugs, there was a highly significant correlation between changes in aortic AIx obtained from the transfer function and radial AIx. That a close relation between aortic and radial AIx should exist is not, as discussed above, unexpected. What is surprising is the simple, approximately linear form of the relation. This means that radial AIx may provide an equally good measure of central pressure wave reflection and has the advantage that it is not dependent on a relatively arbitrary manipulation of the waveform derived from observations made in selected groups of subjects.
In conclusion, our findings question the use of a transfer function to obtain accurate measurements of aortic AIx but suggest that similar information on central pressure wave reflection can be obtained directly from the radial pulse.
Perspectives
Systolic blood pressure and the contour of the pulse waveform differ between central and peripheral arteries. With the recognition that important information relating to arterial wave reflections can be derived from the contour of the aortic pulse, there has been much interest in the estimation of the aortic waveform from more accessible peripheral arteries. A common approach is to use applanation tonometry to obtain the radial or carotid pressure waveform and then to apply a mathematical "transfer function" to synthesize the aortic waveform and hence, to derive the aortic AIx, a measure of arterial pressure wave reflection. The present study compared estimates of aortic systolic blood pressure and the aortic AIx obtained from carotid and radial pulse waveforms by using a transfer function. Those of aortic systolic blood pressure were within acceptable limits of agreement, but those of aortic AIx showed wide variation, especially in healthy subjects. Despite the complexity of the transfer function, an AIx obtained directly from the radial pulse without use of a transfer function showed a close linear correlation with the aortic AIx (estimated by using the transfer function). These findings suggest that the transfer function is of limited value in estimating central arterial wave reflection from the radial pulse and that the same information might be obtained directly from the radial pulse. Future studies should evaluate the use of measurements obtained directly from the peripheral pulse for risk stratification and to identify patients who may benefit from antihypertensive or other treatment.
Received October 22, 2002; first decision November 12, 2002; accepted January 10, 2003.
| References |
|---|
|
|
|---|
2. Nichols WW, ORourke MF. McDonalds Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles. London, UK: Arnold; 1998.
3. Wilkinson IB, MacCallum H, Rooijmans DF, Murray GD, Cockcroft JR, McKnight JA, Webb DJ. Increased augmentation index and systolic stress in type 1 diabetes mellitus. QJM. 2000; 93: 441448.
4. Wilkinson IB, Prasad K, Hall IR, Thomas A, MacCallum H, Webb DJ, Frenneaux MP, Cockcroft JR. Increased central pulse pressure and augmentation index in subjects with hypercholesterolemia. J Am Coll Cardiol. 2002; 39: 10051011.
5. London GM, Blacher J, Pannier B, Guerin AP, Marchais SJ, Safar ME. Arterial wave reflections and survival in end-stage renal failure. Hypertension. 2001; 38: 434438.
6. Kelly R, Hayward C, Ganis J, Daley J, Avolio A, ORourke MF. Non-invasive registration of the arterial pressure pulse waveform using high-fidelity applanation tonometry. J Vasc Med Biol. 1989; 1: 142149.
7. Karamanoglu M, ORourke MF, Avolio AP, Kelly RP. An analysis of the relationship between central aortic and peripheral upper limb pressure waves in man. Eur Heart J. 1993; 14: 160167.
8. Fetics B, Nevo E, Chen CH, Kass DA. Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometry. IEEE Trans Biomed Eng. 1999; 46: 698706.[CrossRef][Medline] [Order article via Infotrieve]
9. Chen CH, Nevo E, Fetics B, Pak PH, Yin FC, Maughan WL, Kass DA. Estimation of central aortic pressure waveform by mathematical transformation of radial tonometry pressure: validation of generalized transfer function. Circulation. 1997; 95: 18271836.
10. Kelly R, Karamanoglu M, Gibbs H, Avolio A, ORourke M. Noninvasive carotid pressure wave registration as an indicator of ascending aortic pressure. J Vasc Med. 1989; 1: 241247.
11. Chen CH, Ting CT, Nussbacher A, Nevo E, Kass DA, Pak P, Wang SP, Chang MS, Yin FC. Validation of carotid artery tonometry as a means of estimating augmentation index of ascending aortic pressure. Hypertension. 1996; 27: 168175.
12. Pauca AL, ORourke MF, Kon ND. Prospective evaluation of a method for estimating ascending aortic pressure from the radial artery pressure waveform. Hypertension. 2001; 38: 932937.
13. Soderstrom S, Nyberg G, ORourke MF, Sellgren J, Ponten J. Can a clinically useful aortic pressure wave be derived from a radial pressure wave? Br J Anaesth. 2002; 88: 481488.
14. van Bortel LM, Balkestein EJ, Heijden-Spek JJ, Vanmolkot FH, Staessen JA, Kragten JA, Vredeveld JW, Safar ME, Struijker-Boudier HA, Hoeks AP. Non-invasive assessment of local arterial pulse pressure: comparison of applanation tonometry and echo-tracking. J Hypertens. 2001; 19: 10371044.[CrossRef][Medline] [Order article via Infotrieve]
15. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1: 307310.[CrossRef][Medline] [Order article via Infotrieve]
16. Kelly RP, Hayward CS, Avolio AP, ORourke MF. Non-invasive determination of age related changes in the human arterial pulse. Circulation. 1989; 80: 16521659.
17. Mahomed FA. The physiology and clinical use of the sphygmograph. Med Times Gazette. 1872; 1: 6265.
This article has been cited by other articles:
![]() |
Y. J. Jeon, J. U. Kim, H. J. Lee, J. Lee, H. H. Ryu, Y. J. Lee, and J. Y. Kim A Clinical Study of the Pulse Wave Characteristics at the Three Pulse Diagnosis Positions of Chon, Gwan and Cheok Evid. Based Complement. Altern. Med., September 29, 2009; (2009) nep150v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Nakae, S. Matsuo, T. Matsumoto, K. Mitsunami, and M. Horie Augmentation Index and Pulse Wave Velocity as Indicators of Cardiovascular Stiffness Angiology, August 1, 2008; 59(4): 421 - 426. [Abstract] [PDF] |
||||
![]() |
M. P Schneider, C. Delles, A. U Klingbeil, M. Ludwig, R. E Kolloch, M. Krekler, K. O Stumpe, and R. E Schmieder Effect of angiotensin receptor blockade on central haemodynamics in essential hypertension: results of a randomised trial Journal of Renin-Angiotensin-Aldosterone System, March 1, 2008; 9(1): 49 - 56. [Abstract] [PDF] |
||||
![]() |
M. Shimizu and K. Kario Review: Role of the augmentation index in hypertension Therapeutic Advances in Cardiovascular Disease, February 1, 2008; 2(1): 25 - 35. [Abstract] [PDF] |
||||
![]() |
R. A. Payne and D. J. Webb Peripheral Augmentation Index: Shouldering the Central Pressure Load Hypertension, January 1, 2008; 51(1): 37 - 38. [Full Text] [PDF] |
||||
![]() |
S. Munir, A. Guilcher, T. Kamalesh, B. Clapp, S. Redwood, M. Marber, and P. Chowienczyk Peripheral Augmentation Index Defines the Relationship Between Central and Peripheral Pulse Pressure Hypertension, January 1, 2008; 51(1): 112 - 118. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Payne, D. Isnardi, P. J. D. Andrews, S. R. J. Maxwell, and D. J. Webb Similarity between the suprasystolic wideband external pulse wave and the first derivative of the intra-arterial pulse wave Br. J. Anaesth., November 1, 2007; 99(5): 653 - 661. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. H. Vanmolkot, L. M. Van Bortel, and J. N. de Hoon Altered arterial function in migraine of recent onset Neurology, May 8, 2007; 68(19): 1563 - 1570. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. J. Kullo and A. R. Malik Arterial Ultrasonography and Tonometry as Adjuncts to Cardiovascular Risk Stratification J. Am. Coll. Cardiol., April 3, 2007; 49(13): 1413 - 1426. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Laurent, J. Cockcroft, L. Van Bortel, P. Boutouyrie, C. Giannattasio, D. Hayoz, B. Pannier, C. Vlachopoulos, I. Wilkinson, H. Struijker-Boudier, et al. Expert consensus document on arterial stiffness: methodological issues and clinical applications Eur. Heart J., November 1, 2006; 27(21): 2588 - 2605. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. J. Stok, B. E. Westerhof, and J. M. Karemaker Changes in finger-aorta pressure transfer function during and after exercise J Appl Physiol, October 1, 2006; 101(4): 1207 - 1214. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Verbeke, P. Segers, S. Heireman, R. Vanholder, P. Verdonck, and L. M. Van Bortel Noninvasive Assessment of Local Pulse Pressure: Importance of Brachial-to-Radial Pressure Amplification Hypertension, July 1, 2005; 46(1): 244 - 248. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. A. Mullan, C. N. Ennis, H. J. P. Fee, I. S. Young, and D. R. McCance Protective effects of ascorbic acid on arterial hemodynamics during acute hyperglycemia Am J Physiol Heart Circ Physiol, September 1, 2004; 287(3): H1262 - H1268. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. N. Cohn, A. A. Quyyumi, N. K. Hollenberg, and K. A. Jamerson Surrogate Markers for Cardiovascular Disease: Functional Markers Circulation, June 29, 2004; 109(25_suppl_1): IV-31 - IV-46. [Full Text] [PDF] |
||||
![]() |
M. F. O'Rourke, A. Avolio, A. Qasem, S. C. Millasseau, J. M. Ritter, and P. J. Chowienczyk Clinical Assessment of Wave Reflection * Response: Augmentation Index and the Radial-to-Aortic Transfer Function Hypertension, November 1, 2003; 42 (5): e15 - e16. [Full Text] [PDF] |
||||
![]() |
S. A. Hope, I. T. Meredith, J. D. Cameron, S. C. Millasseau, J. M. Ritter, and P. J. Chowienczyk Is There Any Advantage to Using an Arterial Transfer Function? * Response: Aortic Augmentation Index and Radial-to-Aortic Transfer Function Hypertension, September 1, 2003; e7(3): . [Full Text] [PDF] |
||||
![]() |
G. E. McVeigh Pulse Waveform Analysis and Arterial Wall Properties Hypertension, May 1, 2003; 41(5): 1010 - 1011. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Hypertension Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |