Donate Help Contact The AHA Sign In Home
American Heart Association
Hypertension
Search: search_blue_button Advanced Search
Hypertension. 2003;41:e9-
Published online before print April 21, 2003, doi: 10.1161/01.HYP.0000070026.52722.76
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
41/6/e9    most recent
01.HYP.0000070026.52722.76v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by de Simone, G.
Right arrow Articles by Pennell, D. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by de Simone, G.
Right arrow Articles by Pennell, D. J.
Related Collections
Right arrow Hypertrophy
Right arrow CT and MRI
Right arrow Echocardiography

(Hypertension. 2003;41:e9.)
© 2003 American Heart Association, Inc.


Letters to the Editor

What Is Bright Is Not Always Gold

Giovanni de Simone; Richard B. Devereux

Department of Clinical and Experimental Medicine, Federico II University Hospital, Naples, Italy, Division of Cardiology, The New York Presbyterian Hospital, Weill Medical College of Cornell University, New York, NY

To the Editor:

The recent paper from Myerson et al1 on the "reliability" of formulae for calculation of echocardiographic left ventricular (LV) mass needs to be carefully considered.

Quantitative echocardiography presents problems of reproducibility, but less than expected from initial studies performed without 2D guidance. There is evidence that the test-retest "error" of echocardiography is similar to the "normal" error affecting many biological determinations.2,3 In particular, in the Reliability of M-mode Echocardiographic Studies (RES) trial,2 using M-mode tracings from several echocardiography laboratories of different experience, test-retest variability was in the range of 10% to 15% (90% confidence interval). Most error is related to image quality and temporal drift, but, as previously reported,2,3 error can be minimized by both appropriate standardization of the examination and in-laboratory self-learning procedures. Reliability can be high, with a coefficient of test-retest concordance of 87% for identification of prognostically adverse LV hypertrophy (LVH).2 Thus, together with accuracy (proved by necropsy validation4–6), reproducibility of echocardiographic LV mass2,3 is better than pessimistically stated by Myerson et al.1,7 Moreover, Daniels et al5 showed that the geometric formula used for echocardiographic LV mass determination predicts necropsy LV weight accurately from linear LV wall thickness and chamber dimension measurements.

Myerson et al1 tried to demonstrate the limited "reliability" of echocardiographic LV mass using a magnetic resonance imaging (MRI) reference standard that had several limitations, including use of a machine with low magnetic field (0.5 Tesla), with low spatial resolution and 1 cm slice thickness, whereas use of higher magnetic fields (1.5 Tesla on average, but up to 3 Tesla) are strongly recommended for cardiac cine-MRI.8 In Myerson’s study,1 measurements were taken "just basal to the papillary muscles," which, with a 1 cm slice thickness, may not have been standardized exactly at the papillary muscle tips. These pitfalls might be especially aggravated by the intrinsic difficulties of MRI in detecting (and reproducing) the largest LV minor axis, because of great slice thickness and difficulty in aligning acquisitions in near-perfect orthogonal planes with LV structures. These limitations can be offset by echocardiography, using orthogonal images or, especially, long axis with precise anatomic markers.

If these limitations are taken into account, the poor result shown in Figure 3 of the paper1 can be explained by higher tool-related technical variability in obtaining a single tomographic slice at the LV minor axis, as compared with using Simpson’s rule. The test-retest variability reported by Myerson et al1 should in fact be compared with analogous tests performed in Prospective Randomized Enalapril Study Evaluating Regression of Ventricular Enlargement (PRESERVE; mostly 2D) and RES (only M-mode). In the RES, for the same 95% confidence interval, test-retest variability was ±35 g from echocardiograms performed in 16 different centers, compared with the higher values (±45 to 55 g) from applying echocardiographic formulae to single-slice MRI data.

It should be emphasized also that what is clinically important is the ability to detect prognostically adverse values and significant changes,9 both of which are accomplished by echocardiography.2,3 Technical variability is <15% at the 90% confidence limit, similar to other biological determinations. Of fundamental importance, we have performed a necropsy comparison study that showed no systematic over- or underestimation of LV mass,4 and numerous investigators have demonstrated the prognostic relevance of echocardiographic LV mass.

Based on results of the PRESERVE study, groups of 41 patients per treatment arm are sufficient to provide statistical power of 90% at an {alpha} error level of 1% to detect a between-group difference of at least 10 g/m2 in LV mass. Thus, echocardiographic LV mass remains an excellent bioassay for clinical trials. Although higher magnetic fields and temporal resolution are improving the performance of cardiac MRI, advances in harmonic imaging and in 3D ultrasound are producing parallel improvements in the ability of echocardiography to image the LV and calculate its mass.

References

1. Myerson SG, Montgomery HE, World MJ, Pennell DJ. Left ventricular mass: reliability of M-mode and 2-dimensional echocardiographic formulas. Hypertension. 2002; 40: 673–678.[Abstract/Free Full Text]

2. de Simone G, Muiesan ML, Ganau A, Longhini C, Verdecchia P, Palmieri V, Agabiti-Rosei E, Mancia G. Reliability and limitations of echocardiographic measurement of left ventricular mass for risk stratification and follow-up in single patients: the RES trial. Working Group on Heart and Hypertension of the Italian Society of Hypertension. Reliability of M-mode Echocardiographic Studies. J Hypertens. 1999; 17 (pt 2): 1955–1963.[CrossRef][Medline] [Order article via Infotrieve]

3. Palmieri V, Dahlof B, DeQuattro V, Sharpe N, Bella JN, de Simone G, Paranicas M, Fishman D, Devereux RB. Reliability of echocardiographic assessment of left ventricular structure and function: the PRESERVE study. Prospective Randomized Study Evaluating Regression of Ventricular Enlargement (see comments). J Am Coll Cardiol. 1999; 34: 1625–1632.[Abstract/Free Full Text]

4. Devereux RB, Alonso DR, Lutas EM, Gottlieb GJ, Campo E, Sachs I, Reichek N. Echocardiographic assessment of left ventricular hypertrophy: comparison to necropsy findings. Am J Cardiol. 1986; 57: 450–458.[CrossRef][Medline] [Order article via Infotrieve]

5. Daniels SR, Meyer RA, Liang YC, Bove KE. Echocardiographically determined left ventricular mass index in normal children, adolescents and young adults. J Am Coll Cardiol. 1988; 12: 703–708.[Abstract]

6. de Simone G, Verdecchia P, Schillaci G, Devereux RB. Clinical impact of various geometric models for calculation of echocardiographic left ventricular mass (see comments). J Hypertens. 1998; 16: 1207–1214.[Medline] [Order article via Infotrieve]

7. Myerson SG, Bellenger NG, Pennell DJ. Assessment of left ventricular mass by cardiovascular magnetic resonance. Hypertension. 2002; 39: 750–755.[Abstract/Free Full Text]

8. Noeske R, Seifert F, Rhein KH, Rinneberg H. Human cardiac imaging at 3 T using phased array coils. Magn Reson Med. 2000; 44: 978–982.[CrossRef][Medline] [Order article via Infotrieve]

9. Verdecchia P, Schillaci G, Borgioni C, Ciucci A, Gattobigio R, Zampi I, Reboldi G, Porcellati C. Prognostic value of serial changes in left ventricular mass in essential hypertension. Circulation. 1998; 97: 48–54.[Abstract/Free Full Text]

Response: What Is Old Is Not Always Best

Saul G. Myerson; Dudley J. Pennell

Cardiovascular Magnetic Resonance Unit, Royal Brompton Hospital, London, UK

The points raised by de Simone and Devereux with regard to our paper,1 fall into 2 areas: (1) criticism of the cardiovascular magnetic resonance (CMR) technique and (2) other studies of the reproducibility of echocardiographically determined left ventricular mass.

(1) The spatial resolution of CMR is good, even at 0.5 Tesla (T). Higher field strengths yield better signal-to-noise ratios, but there is no recommendation to use higher field strengths for cine CMR;2 1.5 T has become routine through availability, and more advanced sequences are programmed for this field strength. However, because improved spatial resolution and signal at higher field strength would only improve the accuracy and reproducibility of the CMR measurements, the argument that the CMR technique was limited is ultimately self-defeating, because, even at lower field strength, CMR exposes the limitations of echocardiography.

The correspondents are incorrect in stating that CMR has difficulty in detecting the true axes of the heart. On the contrary, the strength of CMR is the excellent visualization of any plane within the heart, which makes identification of the true long and short axes more accurate. In particular, the short axis is identified and placed exactly perpendicular to the long axes. Echocardiography has particular problems identifying the true short axes because of the limitations of acoustic access and the inability to determine whether the imaging plane is perpendicular to the long axes.

(2) The reproducibility of echocardiographically determined left ventricular (LV) mass in other studies has been covered elsewhere, including in our recent review article,3 and it was not the purpose of our paper to review this again. The correspondents quote 2 papers suggesting 95% confidence intervals of ±35 g for reproducibility but ignore other papers with wider confidence intervals, such as that of Gottdiener et al,4 which also used 2D-guided M-mode measurements and a single operator and suggested values of ±59 g. Our study did not assess test-retest reliability, because there was a change in LV mass with exercise which varied among individuals. We examined the scatter of values (and confidence limits) of the measured change, and it is not surprising if this was greater than the scatter of 2 measurements of an unchanged LV mass.

Echocardiography has played a fundamental role in assessing LV hypertrophy and determining its prognostic importance. Clearly, it would be possible to continue to examine LV mass with echocardiography in the future. The greater variability of measurements, however, requires more subjects than would be required with CMR, which is why the seminal studies on LV hypertrophy and prognosis5–7 needed their large number of subjects to reach their conclusions. Certainly when such large numbers are either not available (gene studies with low allele frequencies8 or multiple subgroups,9 for example) or not desirable (cost), then CMR has a significant advantage.

References

1. Myerson SG, Montgomery HE, World MJ, Pennell DJ. Left ventricular mass: reliability of M-mode and 2-dimensional echocardiographic formulas. Hypertension. 2002; 40: 673–678.[Abstract/Free Full Text]

2. Bellenger N, Pennell D. Ventricular function. In: Manning W, Pennell DJ, eds. Magnetic Resonance Imaging in Clinical Cardiology. New York, NY: Churchill Livingstone; 2002.

3. Myerson SG, Bellenger NG, Pennell DJ. The assessment of left ventricular mass by cardiovascular magnetic resonance. Hypertension. 2002; 39: 750–755.[Abstract/Free Full Text]

4. Gottdiener J, Livengood S, Meyer P, Chase G. Should echocardiography be performed to assess effects of antihypertensive therapy? Test-retest reliability of echocardiography for measurement of left ventricular mass and function. J Am Coll Cardiol. 1995; 25: 424–430.[Abstract]

5. Levy D, Garrison R, Savage D, Kannel W, Castelli W. Prognostic implications of echocardiographically determined left ventricular mass in the Framingham Heart Study. N Engl J Med. 1990; 322: 1561–1566.[Abstract]

6. Sullivan J, Vander Zwaag R, el-Zeky F, Ramanathan K, Mirvis D. Left ventricular hypertrophy: effect on survival. J Am Coll Cardiol. 1993; 22: 508–513.[Abstract]

7. Ghali J, Liao Y, Simmons B, Castaner A, Cao G, Cooper R. The prognostic role of left ventricular hypertrophy in patients with or without coronary artery disease. Ann Intern Med. 1992; 117: 831–836.[Abstract/Free Full Text]

8. Myerson S, Montgomery H, Whittingham M, Jubb M, World M, Humphries S, Pennell D. Left ventricular hypertrophy with exercise and the angiotensin converting enzyme gene I/D polymorphism: a randomised controlled trial with losartan. Circulation. 2001; 103: 226–230.[Abstract/Free Full Text]

9. Brull D, Dhamrait S, Myerson S, Erdmann J, Regitz-Zagrosek V, World M, Pennell D, Humphries S, Montgomery H. Bradykinin B2BKR receptor polymorphism and left-ventricular growth response. Lancet. 2001; 358: 1155–1156.[CrossRef][Medline] [Order article via Infotrieve]





This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
41/6/e9    most recent
01.HYP.0000070026.52722.76v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by de Simone, G.
Right arrow Articles by Pennell, D. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by de Simone, G.
Right arrow Articles by Pennell, D. J.
Related Collections
Right arrow Hypertrophy
Right arrow CT and MRI
Right arrow Echocardiography