| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
(Hypertension. 2003;42:150.)
© 2003 American Heart Association, Inc.
Scientific Contributions |
From the Department of Medicine, Upstate Medical University, State University of New York at Syracuse (H.S., D.S.S., R.J.C.); the Centre Hospitalier (G.M.L.), F.H. Manhes, Fleury-Merogis, France; and the Department of Internal Medicine, Broussais Hospital (M.E.S.), Paris, France.
Correspondence to Harold Smulyan, MD, Department of Medicine, Upstate Medical University, 750 E Adams St, Syracuse, NY 13210. E-mail smulyanh{at}upstate.edu
| Abstract |
|---|
|
|
|---|
Key Words: aorta arterial pressure Fourier analysis blood pressure determination
| Introduction |
|---|
|
|
|---|
Aortic pressures can now only be measured accurately by invasive catheterization, a method unsuitable for widespread clinical use. ORourke and colleagues25 have proposed a solution involving transformation of the applanated and calibrated radial-artery pulse trace to the aortic pulse trace by Fourier analysis and transfer functions. These and subsequent studies have validated the technique by calibrating radial pulses with directly recorded, intra-arterial radial or brachial pressures.68 The aim of the present study then was to evaluate the accuracy of aortic pressures, calculated from applanated radial-arterial pulses, by comparing them to directly recorded aortic pressures. These noninvasive radial pulses were calibrated by using routine, oscillometric measurements of BA pressures.
| Methods |
|---|
|
|
|---|
A 6F micromanometer-tipped catheter (model SPC 350, Millar Instruments) was adjusted to baseline, calibrated under saline at the catheterization table top, and then inserted through a femoral sheath into the descending aorta under fluoroscopic control. When possible, the catheter was advanced (25 patients) into the ascending aorta, where pressures were recorded, and the catheter was then withdrawn into the proximal descending aorta. Here, systolic (SBP), diastolic (DBP), and mean (MBP) blood pressures were recorded and averaged over several respiratory cycles. MBP was obtained by electronic damping. The frequency-amplitude performance of the catheterization laboratory amplifiers was tested by using both electrical and pressure sine-wave generators. At 7.5 and 10.0 Hz, there was 12% and 20% loss of amplitude, respectively, with a 30% amplitude loss (-3 dB) at 14 Hz. The pressure amplifiers were therefore unsuitable for accurate measurement of augmentation index and ejection times. The mean±SD heart rate for the group was 68±12.2 beats per minute. Ejection fraction (EF) data were available for 38 of the 50 patients by contrast ventriculography at the time of catheterization, by echocardiography, or by radionuclide blood-pool scanning shortly before or after catheterization.
During the direct recording of descending aortic pressure, the left radial-arterial pulse traces were obtained by applanation with a probe (model SPT 301, Millar Instruments). All radial pulses were recorded by the same investigator (H.S.). The mean±SD height of the applanated pulses was 159.5±56.3 mV. The average variability of the radial pulses in systole (average of the differences between each systolic height and the mean systolic height relative to the mean systolic height) was 5.14±2.6%, and in diastole (average of the differences between each diastolic height and the mean diastolic height relative to the mean diastolic height), it was 5.36±3.1%. Thus, in approximately one half of the studies, the variability in systole and diastole was >5%.
The output of the applanation probe was inspected on screen, and when satisfactory stable pulses were observed, 10 seconds of data were fed into the Sphygmocor system (PWV Medical). The radial pulses were calibrated by using the SBPs and DBPs of an oscillometric BA cuff system (Colin Medical Instruments Corp) measured immediately after the radial pulses in the same arm were recorded. MBP was calculated from the radial-artery pulses by integration. The oscillometric blood pressure system was itself calibrated by using a calibration unit (Cuff Link, DNI Nevada Inc).
Correlation coefficients between the various measured and calculated parameters were obtained by univariate regression. Multivariable regression was used to detect the influence of nonpressure pulse factors that might improve the prediction of direct aortic values from the calculated ones.
In a small substudy, the oscillometric cuff pressure technique was compared with the more traditional auscultatory method by using a mercury-calibrated aneroid manometer and phases I and V of the Korotkoff sounds. This comparison was carried out with single sequential oscillometric and auscultatory blood pressure measurements in the left arm of 50 similarly selected patients.
| Results |
|---|
|
|
|---|
|
With regard to SBP, the average difference between the Millar SBP and the cuff SBP was, as expected, wide (137.2 vs 148.1 mm Hg), and the SD of the differences was large (12.7 mm Hg). By contrast, the average measured and calculated values for SBP were quite close (137.2 vs 135.7 mm Hg), but the SD of the differences was approximately the same. The "predicted" values in Table 1 are those obtained from multiple regression (vide infra). Figure 1A is a plot of the measured versus calculated SBP and shows a slope that is close to 1.0, an intercept close to zero, and a correlation coefficient (r) of +0.89. Figure 1B is a Bland-Altman plot of the same data showing the measured-calculated differences versus the directly measured values.
|
The average difference between measured and calculated DBPs shown in Table 1 was large (74.4 vs 84.8 mm Hg) because of overestimation of DBP by the cuff (83.3 mm Hg). The scatter of both calculated and cuff DBP differences was also large. The plot of the measured versus calculated DBP values is shown in Figure 2A. Here, the slope and intercept are distant from 1.0 and zero, respectively, and the r value was only +0.59. Figure 2B displays the measured-calculated DBP differences versus the measured DBP values in a Bland-Altman plot. The overestimation of aortic DBP by the Sphygmocor device accounts for the underestimation of its PP by approximately the same amount (11.5 mm Hg). Figure 3A illustrates the measured versus calculated PP data with some improvement in the slope and intercept when compared with the DBP plot, but the SD of the PP differences remained high (13.6 mm Hg). Figure 3B is the corresponding Bland-Altman plot. The upward slope of this plot is due to progressive overestimation by the cuff DBP at higher values (Fig 2A). As shown in Table 1, the average cuff SBP was higher than the measured SBP, but the slope of the relation was near 1.0 (data not shown).
|
|
As expected, comparisons of ascending and descending aortic SBP, DBP, and PP values were quite close, with r values of +0.97, +0.92, and +0.98, respectively. All of the univariate analyses were repeated for men and women separately, and no differences in the results were observed. Differences between measurements were also analyzed for age, but no effect of age on the results was found.
Multivariable Analyses
The aortic SBP and PP must also be determined by factors unaccounted for by the Sphygmocor device alone. Separate univariate analyses identified age, height, heart rate, and EF as variables that could contribute to the accuracy of the Sphygmocor-calculated pressures. Measured EF, however, was only available for 38 cases. The mean EF for the available 38 cases was used in place of the missing 12 values to avoid loss of the other independent variables in those cases without a measured EF. With regard to each dependent variable, SBP was significantly influenced by age and height, DBP by age and heart rate, and PP by age and EF (P=0.08 for 38 cases). The values for the nonstandardized regression coefficients for each of the independent variables (including the Sphygmocor values), the intercept for the estimating equations, and the partial probability values are shown in Table 2. From the estimating equations provided by multiple regression, each of the pressures was recalculated ("predicted" values) and compared with the data obtained from direct aortic measurements. The mean predicted values and their differences from the Millar catheter-generated values are shown in Table 1, where they also may be compared with those from the Sphymocor. It is apparent that use of multiple regression brings the mean differences to near zero and slightly reduces the scatter. No predicted cuff DBP values are shown because the cuff DBP was assumed by the Sphygmocor device to be the same as aortic DBP. Addition of other independent variables to the Sphygmocor data alone can also be appreciated by inspection of Table 1. Although mean differences are brought to near zero, the SDs are only slightly improved. Multiple regression necessarily brings each of the univariate regression-line slopes to near identity and improves the coefficient of determination (R2) and r values (Table 3). Statistical comparison of the R2 values for the univariate relations between measured and calculated pressures with those obtained from multiple regression (F test) revealed that the multivariable analysis significantly improved all R2s (P<0.01). Table 3 also compares the r and R2 values for the calculated versus cuff blood pressure values in both univariate and multivariate analyses. Except for the univariate R2 and r for PP, there is little improvement of the Sphygmocor data over that of the cuff. Interactive effects of the independent variables used in multiple regression were sought, and none were found.
|
|
Figures 4A and 4B show the direct and Bland-Altman plots comparing the auscultatory with the oscillometric method of measuring SBP, and Figures 5A and 5B show similar plots for DBP. Had the study been done with the auscultatory method, the cuff SBP would, on average, have been 7.1 mm Hg lower, but there would have been no difference in average DBP.
|
|
| Discussion |
|---|
|
|
|---|
Lasance et al,4 Karamanoglu et al,5 and Pauca et al8 have presented and Nichols and ORourke9 have championed such a method by using a calibrated, applanated, radial-artery pulse trace and transforming it into an aortic pulse trace by Fourier analysis and transfer functions. The original descriptions validated the method in 14 patients with near-simultaneous calculated and directly measured radial and aortic pulses. To this was added a number of other cases drawn from the literature that displayed directly recorded (not applanated) brachial or radial pulses and aortic pulses. Several other reports have also validated the transfer-function technique, but each has calibrated the radial pulses with direct recordings of arterial or aortic pulses.68 If such a system is to be acceptable for widespread clinical use, it must be validated without the need for direct brachial or radial-artery recordings as calibrations. The present study approached this problem by using the Sphygmocor device to calculate the aortic pulses from applanated radial pulses but calibrated the radial pulses by using a standard oscillometric cuff unit. Our data show the expected overestimation of aortic SBP by the cuff method (reduced by aging) and its average correction by the Sphygmocor device. Unfortunately, the correction is associated with large scatter, too large to meet the British or US standards for the substitution of 1 technique for the other.12,13 The cuff DBPs were unexpectedly high when compared with the directly recorded aortic DBPs. Inaccuracy of the cuff pressures is not surprising, because previous comparisons of pressures obtained with the time-honored, standard mercury device with intra-arterial values have shown mean differences for SBP ranging from 0.9 to 12.3, with SDs from 1.3 to 13.0 mm Hg. Mean differences for DBP ranged from 8.3 to 18, with SDs from 1.1 to 9.3 mm Hg.13 Some comparisons of the auscultatory versus oscillometric methods have been good,14 some poor (especially DBP1519), some with oscillometric pressures too high,20 and some too low21 in a variety of clinical populations. Only a single, early study measured oscillometric arm pressures along with directly recorded aortic pressures and found them comparable.22 We chose the oscillometric method because of its increasing acceptance in clinical work, avoidance of interpretive errors (digit preferences), and ease of use in the catheterization laboratory. The substudy of auscultatory versus oscillometric BA pressures showed that the substitution of auscultatory for oscillometric BA pressures would have reduced the good relation between measured and calculated SBP, but the discrepancy between oscillometric and aortic DBPs would, on average, have been unchanged.
In the present study, oscillometric cuff DBPs were higher than the aortic DBPs. This technical difference accounted for the calculated underestimation of PP and the overestimation of MBP. On the other hand, the cuff method also overestimated SBP and DBP by almost the same amount, resulting in a much closer prediction of aortic PP (difference of 2.5 mm Hg).
There are many possible explanations for the disparities and scatter in the results. Radial pulses were unstable and poorly recorded. All recordings were made by a single individual (H.S.), and pulse heights were those recommended in the Sphygmocor manual. Approximately one half of the recordings had systolic and diastolic variability >5%, levels higher than recommended in the instrument manual. When the SBP data for the 24 patients whose systolic and diastolic variability was 5% or less were analyzed separately, there was no difference from the entire group in the r value (+0.87) or in the SD of the differences (±13.2 mm Hg.). This makes technical variability in radial-pulse recording an unlikely explanation. Another possible source of variability is the spontaneous oscillations in diameter and distensibility known to occur in the human radial artery.23 Catheterization patients were unstable. Many of the patients had been sedated and some were undoubtedly anxious, resulting in respiratory variations in aortic pressures. These variations should have been smoothed out by averaging the pressures across respiratory cycles but could have affected the stability of the radial-artery pulses. The cuff pressures and radial and aortic pulses were not exactly simultaneous. Because we wished to have the radial pulses and cuff pressures in the same arm, simultaneity was not possible. The radial pulses were recorded simultaneously with those of aortic pressures, but there was no beat-to-beat correspondence. It was assumed that BA pressure is the same as the radial-artery pressure. Although they cannot be exactly the same, the differences should be small. Errors were made in the measurement of BA pressure by the oscillometric technique. This is probably the largest source of measurement error in the study. It is a single calibration pressure for a series of radial-artery pulses reduced to an average pulse. Although manufacturers algorithms vary, the most accurate measurement by the oscillometric method is the MBP, which is selected as the cuff pressure when the oscillations are at their maximum. The use of averaged, multiple-cuff MBPs for calibration purposes, in settings more stable than the catheterization laboratory, would likely increase the accuracy of these values.
It has been recently suggested24 that cuff pressures performed about as well (or better, for SBP) than the Sphygmocor device in the prediction of aortic pressures. In the present study, the SBP was better predicted by the Sphygmocor device, but the discrepancy between the DBPs was about the same. In our study, however, the oscillometric cuff was independently calibrated, and the aortic pressures were recorded with a transducer-tip rather than a fluid-filled catheter. These methodological differences could account in part for the differences in the results.
Obviously, the generalized transfer function cannot account for all of the factors that influence the way in which the aortic pulse is transmitted to the radial artery. When age, height, heart rate, and EF were incorporated into a multivariable analysis, there was a significant improvement (F test; P<0.01) in the correlation of calculated pressures with direct aortic measurements, a reduction in the scatter, and a mathematically obligatory elimination of the mean differences. When the same independent variables were incorporated into the analyses for cuff pressures, similar improvements in the differences, correlations, and scatter occurred, PP being the most improved.
Perspectives
An accurate yet noninvasive means for measuring central aortic pressure would be an important clinical advance over that provided by BA pressure. The Sphygmocor device has the potential to fill this need. This study shows that the Sphygmocor method is limited by the variability and inaccuracy of the cuff blood pressure required for radial-artery calibration. Comparison of calculated and measured aortic pressures might be improved if MBP only were used for calibration and multiple cuff measurements over a short period of time had been available and averaged. Although further validation studies requiring direct aortic pressures in a variety of clinical situations will be difficult to obtain, they should be pursued, because the clinical benefits would be great if the device can realize its potential.
| Acknowledgments |
|---|
Received April 16, 2003; first decision May 9, 2003; accepted June 12, 2003.
| References |
|---|
|
|
|---|
2. Kelly R, Hayward C, Ganis J, Daley J, Avolio A, ORourke M. Noninvasive registration of the arterial pressure pulse waveform using high-fidelity applanation tonometry. J Vasc Med Biol. 1989; 1: 142149.
3. ORourke MF. Influence of ventricular ejection on the relationship between central aortic and brachial pressure pulse in man. Cardiovasc Res. 1970; 4: 291300.
4. Lasance HAJ, Wesseling KH, Ascoop CA. Peripheral pulse contour analysis in determining stroke volume. Progress Report 5. Utrecht, Netherlands: Institute of Medical Physics, 1976: 5962.
5. Karamanoglu M, ORourke MF, Avolio AP, Kelly RP. An analysis of the relationship between central aortic and peripheral upper limb pressure waves in man. Eur Heart J. 1993; 14: 160167.
6. Chen C-H, Nevo E, Fetics B, Pak PH, Yin FCP, Maughan L, Kass DA. Estimation of central aortic pressure waveform by mathematical transformation of radial tonometry pressure: validation of generalized transfer function. Circulation. 1997; 95: 18271836.
7. Fetics B, Nevo E, Chen C-H, Kass DA. Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometry. IEEE Trans Biomed Eng. 1999; 46: 698706.[CrossRef][Medline] [Order article via Infotrieve]
8. Pauca AL, ORourke MF, Kon ND. Prospective evaluation of a method for estimating ascending aortic pressure from the radial artery pressure waveform. Hypertension. 2001; 38: 932937.
9. Nichols WW, ORourke MF. Sphygmocardiography: ascending aortic pressure waves. In: McDonalds Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles. 4th ed. London/Sydney/Auckland: Arnold; 1998: 187, 457476.
10. Cohn JN. Blood pressure measurement in shock: mechanism of inaccuracy in auscultatory and palpatory methods. JAMA. 1967; 199: 972976.
11. ORourke MF, Pauca A, Jiang X-J. Pulse wave analysis. Br J Clin Pharmacol. 2001; 51: 507522.[CrossRef][Medline] [Order article via Infotrieve]
12. OBrien E, Petrie J, Littler W, deSwiet M, Padfield PL, OMalley K, Jamieson M, Altman D, Bland M, Atkins N. The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens. 1990; 8: 607619.[Medline] [Order article via Infotrieve]
13. American National Standard for Electronic or Automated Sphygmomanometers. ANSI/AAMI SP101992. Arlington, Va: Association for the Advancement of Medical Instrumentation. 1993: 12.
14. Jones CR, Taylor K, Poston L, Shennen AH. Validation of the Welch Allyn "Vital Signs" oscillometric blood pressure monitor. J Hum Hypertens. 2001; 15: 191195.[CrossRef][Medline] [Order article via Infotrieve]
15. Nystrom E, Reid KH, Bennett R, Couture L, Edmonds HL Jr. A comparison of two automated indirect arterial blood pressure meters: with recordings from a radial artery catheter in anesthetized surgical patients. Anesthesiology. 1985; 62: 526530.[CrossRef][Medline] [Order article via Infotrieve]
16. Runcie CJ, Reeve WG, Reidy J, Dougall JR. Blood pressure measurement during transport: a comparison of direct and oscillometric readings in critically ill patients. Anaesthesia. 1990; 45: 659665.[Medline] [Order article via Infotrieve]
17. Venus B, Mathru M, Smith RA, Pham CG. Direct versus indirect blood pressure measurements in critically ill patients. Heart Lung. 1985; 14: 228231.[Medline] [Order article via Infotrieve]
18. Rithalia SVS, Edwards D. Comparison of oscillometric and intra-arterial blood pressure and pulse measurements. J Med Eng Technol. 1994; 18: 179181.[Medline] [Order article via Infotrieve]
19. OBrien E, Mee F, Atkins N, OMalley K. Short report: accuracy of the Dinamap portable monitor, model 8100 determined by the British Hypertension Society Protocol. J Hypertens. 1993; 11: 761763.[CrossRef][Medline] [Order article via Infotrieve]
20. Park MK, Menard SW, Yuan C. Comparison of auscultatory and oscillometric blood pressures. Arch Ped Adolesc Med. 2001; 155: 5053.
21. Bur A, Hirschl MM, Herkner H, Oschatz E, Kofler J, Woisetschlager C, Laggner AN. Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients. Crit Care Med. 2000; 28: 371376.[CrossRef][Medline] [Order article via Infotrieve]
22. Borow KM, Newburger JW. Noninvasive estimation of central aortic pressure using the oscillometric method for analyzing systemic artery pulsatile blood flow: comparative study of indirect systolic, diastolic, and mean brachial artery pressure with simultaneous direct ascending aortic pressure measurements. Am Heart J. 1982; 103: 879886.[CrossRef][Medline] [Order article via Infotrieve]
23. Hayoz D, Tardy Y, Rutschmann B, Mignot JP, Achakri H, Feihl F, Meister JP, Waeber B, Brunner HR. Spontaneous diameter oscillations of the radial artery in humans. Am J Physiol. 1993; 264: H2080H2084.[Medline] [Order article via Infotrieve]
24. Davies JI, Band MM, Pringle P, Ogston S, Struthers AD. Peripheral blood pressure measurement is as good as applanation tonometry at predicting ascending aortic pressure. J Hypertens. 2003; 21: 571576.[CrossRef][Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
J. Andrade, L. Er, A. Ignaszewski, and A. Levin Exploration of Association of 1,25-OH2D3 with Augmentation Index, a Composite Measure of Arterial Stiffness Clin. J. Am. Soc. Nephrol., November 1, 2008; 3(6): 1800 - 1806. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D. Smith and P. J. Levy Review: New techniques for assessment of vascular function Therapeutic Advances in Cardiovascular Disease, October 1, 2008; 2(5): 373 - 385. [Abstract] [PDF] |
||||
![]() |
M. E. Safar and H. Smulyan Central Blood Pressure and Hypertension Hypertension, April 1, 2008; 51(4): 819 - 820. [Full Text] [PDF] |
||||
![]() |
M. Shimizu and K. Kario Review: Role of the augmentation index in hypertension Therapeutic Advances in Cardiovascular Disease, February 1, 2008; 2(1): 25 - 35. [Abstract] [PDF] |
||||
![]() |
E. Agabiti-Rosei, G. Mancia, M. F. O'Rourke, M. J. Roman, M. E. Safar, H. Smulyan, J.-G. Wang, I. B. Wilkinson, B. Williams, and C. Vlachopoulos Central Blood Pressure Measurements and Antihypertensive Therapy: A Consensus Document Hypertension, July 1, 2007; 50(1): 154 - 160. [Full Text] [PDF] |
||||
![]() |
V. Palmieri, R. Pini, and M. Chiara Cavallini Pulsology Reloaded: Commentary on Similar Effects of Treatment on Central and Brachial Blood Pressure in Older Hypertensive Subjects Hypertension, June 1, 2007; 49(6): 1210 - 1212. [Full Text] [PDF] |
||||
![]() |
J. D. Cameron, I. T. Meredith, and S. A. Hope Letter by Cameron et al Regarding Article, "Differential Impact of Blood Pressure-Lowering Drugs on Central Aortic Pressure and Clinical Outcomes: Principal Results of the Conduit Artery Function Evaluation (CAFE) Study" Circulation, October 10, 2006; 114(15): e538 - e538. [Full Text] [PDF] |
||||
![]() |
A. M. Dart, C. D. Gatzka, B. A. Kingwell, K. Willson, J. D. Cameron, Y.-L. Liang, K. L. Berry, L. M.H. Wing, C. M. Reid, P. Ryan, et al. Brachial Blood Pressure But Not Carotid Arterial Waveforms Predict Cardiovascular Events in Elderly Female Hypertensives Hypertension, April 1, 2006; 47(4): 785 - 790. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Oparil and J. L. Izzo Jr Pulsology Rediscovered: Commentary on the Conduit Artery Function Evaluation (CAFE) Study Circulation, March 7, 2006; 113(9): 1162 - 1163. [Full Text] [PDF] |
||||
![]() |
F. Verbeke, P. Segers, S. Heireman, R. Vanholder, P. Verdonck, and L. M. Van Bortel Noninvasive Assessment of Local Pulse Pressure: Importance of Brachial-to-Radial Pressure Amplification Hypertension, July 1, 2005; 46(1): 244 - 248. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Chirinos, J. P. Zambrano, S. Chakko, A. Veerani, A. Schob, H. J. Willens, G. Perez, and A. J. Mendez Aortic Pressure Augmentation Predicts Adverse Cardiovascular Events in Patients With Established Coronary Artery Disease Hypertension, May 1, 2005; 45(5): 980 - 985. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. F. O'Rourke and W. W. Nichols Aortic Diameter, Aortic Stiffness, and Wave Reflection Increase With Age and Isolated Systolic Hypertension Hypertension, April 1, 2005; 45(4): 652 - 658. [Full Text] [PDF] |
||||
![]() |
T. G. Pickering, J. E. Hall, L. J. Appel, B. E. Falkner, J. Graves, M. N. Hill, D. W. Jones, T. Kurtz, S. G. Sheps, and E. J. Roccella Recommendations for Blood Pressure Measurement in Humans and Experimental Animals: Part 1: Blood Pressure Measurement in Humans: A Statement for Professionals From the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research Circulation, February 8, 2005; 111(5): 697 - 716. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. G. Pickering, J. E. Hall, L. J. Appel, B. E. Falkner, J. Graves, M. N. Hill, D. W. Jones, T. Kurtz, S. G. Sheps, and E. J. Roccella Recommendations for Blood Pressure Measurement in Humans and Experimental Animals: Part 1: Blood Pressure Measurement in Humans: A Statement for Professionals From the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research Hypertension, January 1, 2005; 45(1): 142 - 161. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. L. Pauca, N. D. Kon, and M. F. O'Rourke The second peak of the radial artery pressure wave represents aortic systolic pressure in hypertensive and elderly patients Br. J. Anaesth., May 1, 2004; 92(5): 651 - 657. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. T. Schram, R. M.A Henry, R. A.J.M van Dijk, P. J. Kostense, J. M. Dekker, G. Nijpels, R. J. Heine, L. M. Bouter, N. Westerhof, and C. D.A. Stehouwer Increased Central Artery Stiffness in Impaired Glucose Metabolism and Type 2 Diabetes: The Hoorn Study Hypertension, February 1, 2004; 43(2): 176 - 181. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. F. O'Rourke, A. Avolio, A. Qasem, S. C. Millasseau, J. M. Ritter, and P. J. Chowienczyk Clinical Assessment of Wave Reflection * Response: Augmentation Index and the Radial-to-Aortic Transfer Function Hypertension, November 1, 2003; 42 (5): e15 - e16. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Hypertension Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |