(Hypertension. 1995;26:835.)
© 1995 American Heart Association, Inc.
Articles |
From The Blood Pressure Unit, Beaumont Hospital (E.OB., N.A.), Dublin, Ireland, and Klinisch Laboratorium Hypertensie, Inwendige Geneeskunde-Cardiologie (J.S.), UZ Gasthuisberg, Leuven, Belgium.
Abstract The introduction of 24-hour ambulatory blood pressure measurement into clinical practice created a large market for ambulatory blood pressure measurement devices. Forty-three such devices from 31 manufacturers or suppliers are now available to satisfy a market demand that is likely to increase. The aim of this article is to identify the devices available and then to examine critically any validation studies assessing accuracy and performance. Of the 43 devices available 18 have been validated according to the protocols of the Association for the Advancement of Medical Instrumentation (AAMI) or the British Hypertension Society (BHS) in 25 reported studies. In 9 of these studies the protocol was not adhered to, and the results, which are therefore questionable, are noted but not considered further. Fourteen devices were evaluated according to the accuracy criteria of both protocols, and of these 9 fulfilled the requirements. From this review of 43 devices on the market it may be concluded that, at the time of writing, there is published evidence for only 9 devices meeting the generally accepted AAMI and BHS criteria for accuracy and performance; these are the A&D TM-2420 models 6 and 7 and TM-2421, CH-Druck, Nissei ABPM DS-240, Profilomat, QuietTrak, and SpaceLabs SL-90202 and SL-90207.
Key Words: blood pressure monitoring, ambulatory equipment design evaluation studies durable medical equipment
This article has been cited by other articles:
![]() |
T. B. Torbjornsdotter, G. A. Jaremko, and U. B. Berg Nondipping and Its Relation to Glomerulopathy and Hyperfiltration in Adolescents With Type 1 Diabetes Diabetes Care, February 1, 2004; 27(2): 510 - 516. [Abstract] [Full Text] [PDF] |
||||
![]() |
F.M. McKevitt, A. Sivaguru, G.S. Venables, T.J. Cleveland, P.A. Gaines, J.D. Beard, and K.S. Channer Effect of Treatment of Carotid Artery Stenosis on Blood Pressure: A Comparison of Hemodynamic Disturbances After Carotid Endarterectomy and Endovascular Treatment Stroke, November 1, 2003; 34(11): 2576 - 2581. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Simckes, T. Srivastava, and U. S. Alon Ambulatory Blood Pressure Monitoring in Children and Adolescents Clinical Pediatrics, October 1, 2002; 41(8): 549 - 564. [Abstract] [PDF] |
||||
![]() |
J. E. Lewis, E. Boyle, L. Magharious, and M. G. Myers Evaluation of a community-based automated blood pressure measuring device Can. Med. Assoc. J., April 1, 2002; 166(9): 1145 - 1148. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Frauchiger, H. P. Schmid, C. Roedel, P. Moosmann, and D. Staub Comparison of Carotid Arterial Resistive Indices With Intima-Media Thickness as Sonographic Markers of Atherosclerosis Stroke, April 1, 2001; 32(4): 836 - 841. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. O'Brien, B. Waeber, G. Parati, J. Staessen, and M. G Myers Blood pressure measuring devices: recommendations of the European Society of Hypertension BMJ, March 3, 2001; 322(7285): 531 - 536. [Full Text] |
||||
![]() |
J. A Staessen, E. T O'Brien, L. Thijs, and R. H Fagard Modern approaches to blood pressure measurement Occup. Environ. Med., August 1, 2000; 57(8): 510 - 520. [Abstract] [Full Text] |
||||
![]() |
E. Lurbe and J. Redon Assessing ambulatory blood pressure in renal diseases: facts and concerns Nephrol. Dial. Transplant., November 1, 1999; 14(11): 2564 - 2568. [Full Text] [PDF] |
||||
![]() |
J. E. Naschitz, L. Loewenstein, E. Zuckerman, D. Yeshurun, J. A. Staessen, and E. O'Brien Reliability of Self-Measured Blood Pressure for Research Purposes Arch Intern Med, October 25, 1999; 159(19): 2365 - 2366. [Full Text] [PDF] |
||||
![]() |
K. Seppa and P. Sillanaukee Binge Drinking and Ambulatory Blood Pressure Hypertension, January 1, 1999; 33(1): 79 - 82. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Thijs, J. A. Staessen, H. Celis, R. de Gaudemaris, Y. Imai, S. Julius, and R. Fagard Reference Values for Self-recorded Blood Pressure: A Meta-analysis of Summary Data Arch Intern Med, March 9, 1998; 158(5): 481 - 488. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Staessen, L. Bieniaszewski, E. O'Brien, P. Gosse, H. Hayashi, Y. Imai, T. Kawasaki, K. Otsuka, P. Palatini, L. Thijs, et al. Nocturnal Blood Pressure Fall on Ambulatory Monitoring in a Large International Database Hypertension, January 1, 1997; 29(1): 30 - 39. [Abstract] [Full Text] |
||||
![]() |
N. Prasad and C. Isles Fortnightly Review: Ambulatory blood pressure monitoring: a guide for general practitioners BMJ, December 14, 1996; 313(7071): 1535 - 1541. [Abstract] [Full Text] |
||||
|
Hypertension Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |